Thermo Scientific 5028i Instruction Manual page 260

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
Diagnostics
B-20
Model 5028i Instruction Manual
diag volt det a/b
This command reports the diagnostic voltage measurements on the
detector board. The voltage is positive 5.
Send:
diag volt det a
Receive:
diag volt det a 4.9
Send:
diag volt det b
Receive:
diag volt det b 4.9
diag volt iob
This command reports the diagnostic voltage measurements on the I/
O expansion board. The sequence of voltages is: Positive 24, positive
5, positive 3.3, and negative 3.3. Each voltage value is separated by a
space.
Send:
diag volt iob
Receive:
diag volt iob 24.10 4.90 3.20 -3.20
diag volt mb
This command reports the diagnostic voltage measurements on the
motherboard. The sequence of voltages is: Positive 24, positive 15,
positive 5, positive 3.3, and negative 3.3. Each voltage value is separated
by a space.
Send:
diag volt mb
Receive:
diag volt mb 24.10 14.90 4.90 3.20 -3.20
diag volt mib
This command reports the diagnostic voltage measurements on the
measurement interface board. The sequence of voltages is: Positive 24,
positive 15, negative 15, positive 5, and positive 3.3. Each voltage value
is separated by a space.
Send:
diag volt mib
Receive:
diag volt mib 24.98 14.80 -14.90 4.96 3.20
diag volt neph
This command reports the diagnostic voltage measurements on the
nephelometer interface board.
Send:
diag volt neph
Receive:
diag volt neph
flow a/b
This command reports the current sample flow. The following example
reports that the sample flow is 16.69 LPM.
Send:
flow a
Receive:
flow a 16.69 LPM