Calibration
Calibration Procedure
Detector Calibration
Auto Detector Calibration
4-10
Model 5028i Instruction Manual
and press
to confirm the calibration routine has now been
competed.
The Detector Calibration submenu is used to view and set the detector
calibration. The detector calibration screen is visible only when the
instrument is in service mode. For more information on the service
mode, see
"Service
Mode" earlier in the chapter.
Note This adjustment should only be performed by an instrument
service technician. Detector Calibration B settings please refer to
Detector Calibration as following description.
In the Main Menu, choose Service > Detector Calibration or
l
Detector Calibration B.
The Auto Detector Calibration screen allows the user to optimize the
detector performance. This should be done at least once per year and as
often as once per quarter.
In the Main Menu, choose Service > Detector Calibration >
l
Detector Calibration or Detector Calibration B > Auto.