Table B-9. Set Layout Ack Values; Table B-10. Power Up Mode Values - Thermo Scientific 5028i Instruction Manual

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
Communications Configuration
B-54
Model 5028i Instruction Manual
Table B–9. Set Layout Ack Values
Value
Function
0
Do nothing (default)
1
Append "*"
power up mode
This command reports the current power up mode setting to either 0
= local/unlocked or 1 = remote/locked, according to
default value is 0; power up in local/unlocked mode. The following
example shows that the instrument is configured to power up in the
remote/locked mode.
Send:
power up mode
Receive:
power up mode 1
set power up mode value
This command is used to configure the instrument to power up in the
local/unlocked mode (value= 0) or the remote/locked mode (value = 1),
according to
Table
B–10.
If the instrument is set to power up in the local/remote mode, the
keypad can be used to make changes via the front panel. If the
instrument is set to power up in the remote/locked mode, changes
can not be made from the front panel. The following example sets the
instrument to power up in remote/locked mode.
Send:
set power up mode 1
Receive:
set power up mode 1 ok
Table B–10. Power Up Mode Values
Value
Power up Mode
0
Local/Unlocked Mode (default)
1
Remote/Locked Mode
program no
This command reports the analyzer's model information and program
version number.
Send:
program no
Receive:
program no iSeries 5028i 00.05.37.093
tz
This command reports the "tz" timezone string for the NTP server. See
"Network Time Protocol Server" in the "Communications Settings"
Table
B–10.The
Thermo Fisher Scientific