C-Link Protocol Commands
Calibration
B-36
Model 5028i Instruction Manual
high volt a/b
This command reports the current high voltage for the detector. The
following example reports that the high voltage is 1400 V.
Send:
high volt a
Receive:
high volt a 1400
Send:
high volt b
Receive:
high volt b 1400
set high volt a/b value
This command sets the high voltage for the detector to value, where
value is a floating-point representation of the high voltage. The
following example sets the high voltage to 1445 V.
Send:
set high volt a 1445
Receive:
set high volt a 1445 ok
Send:
set high volt b 1445
Receive:
set high volt b 1445 ok
mass coef a/b
This command reports the current mass coefficient. The following
example reports that the mass coefficient is 7100.
Send:
mass coef a
Receive:
mass coef a 7100
Send:
mass coef b
Receive:
mass coef b 7100
set mass coef a/b value
This command sets the mass coefficient to user-defined values to value,
where value is a floating-point representation of the coefficient. The
following example sets the mass coefficient to 7000.
Send:
set mass coef a 7000
Receive:
set mass coef a 7000 ok
Send:
set mass coef b 7000
Receive:
set mass coef b 7000 ok
thermal mass coef a/b
This command reports the current thermal mass coefficient. The
following example reports that the thermal mass coefficient is 19.0000.
Send:
thermal mass coef a
Receive:
thermal mass coef a 19.0000
Send:
thermal mass coef b
Receive:
thermal mass coef b 19.0000