This command reports whether temperature compensation is for actual
or standard conditions. The following example shows the temperature
compensation is standard.
Send:
temp comp
Receive:
temp comp std
set temp comp selection
selection = | act | std |
These commands turn the temperature compensation to either actual
or standard conditions. The following example turns the temperature
compensation to actual.
Send:
set temp comp act
Receive:
set temp comp act ok
temp std a/b
This command reports the standard temperature. The following example
reports that the standard temperature is 18 °C.
Send:
temp std a
Receive:
temp std a 18 degC
Send:
temp std b
Receive:
temp std b 18 degC
set temp std a/b value
This command sets the standard temperature to value, where value is a
floating-point number representation of standard temperature between
0 to 25 °C. The following example sets the standard temperature to 15
°C.
Send:
set temp std a 15
Receive:
set temp std a 15 ok
Send:
set temp std b 15
Receive:
set temp std b 15 ok
temp th a/b
This command reports the heater temperature threshold for the sample.
The following example reports that the temperature threshold is 30 °C.
Send:
temp th a
Receive:
temp th a 30 degC
Send:
temp th b
Receive:
temp th b 30 degC
C-Link Protocol Commands
Measurement Configuration
Model 5028i Instruction Manual
B-45