C-Link Protocol Commands
Alarms
B-14
Model 5028i Instruction Manual
Send:
set alarm beta count max 20000
Receive:
set alarm beta count max 20000 ok
alarm board temp min
alarm board temp max
These commands report the current board temperature alarm minimum
and maximum value settings. The following example reports that the
board temperature alarm minimum value is 0 °C.
Send:
alarm board temp min
Receive:
alarm board temp min 0.0 degC
set alarm board temp min value
set alarm board temp max value
These commands set the board temperature alarm minimum and
maximum values to value, where value is a floating-point number
representing board temperature alarm limits in °C. The following
example sets the board temperature alarm maximum value to 38 °C.
Send:
set alarm board temp max 38.0
Receive:
set alarm board temp max 38.0 ok
alarm conc avg pm min a/b
alarm conc avg pm max a/b
These commands report the current 24-hour average PM concentration
alarm minimum and maximum values setting. The following example
reports that the average PM concentration minimum is 100 mg/m
Send:
alarm conc avg pm min a
Receive:
alarm conc avg pm min a 1.000E+02 mg/m3
Send:
alarm conc avg pm min b
Receive:
alarm conc avg pm min b 1.000E+02 mg/m3
set alarm conc avg pm min a/b value
set alarm conc avg pm max a/b value
These commands set the 24-hour average PM concentration alarm
minimum and maximum values to value, where value is a floating-
point representation of the concentration alarm limits. Values must be
in the units that are currently set for use. The following example sets the
average PM concentration alarm maximum value to 10000 mg/m
Send:
set alarm conc avg pm max a 10000
Receive:
set alarm conc avg pm max a 10000 ok
Send:
set alarm conc avg pm max b 10000
Receive:
set alarm conc avg pm max b 10000 ok