C-Link Protocol Commands
Calibration
B-32
Model 5028i Instruction Manual
baro span
This command reports the current barometric pressure span recorded
at the time of calibration. The following example reports that the
barometric span is 1.0.
Send:
baro span
Receive:
baro span 1.0000
set baro span value
This command requires to manually specifying the span value for the
barometric pressure span,where value is a floating-point representation
of barometric pressure span. The following example successfully sets the
barometric span to 1.023.
Send:
set baro span 1.023
Receive:
set baro span 1.023 ok
beta th a/b
This command reports the current reference beta threshold for the
detector. The following example reports that the beta threshold is 440.
Send:
beta th a
Receive:
beta th a 440
Send:
beta th b
Receive:
beta th b 440
set beta th a/b value
This command sets the reference beta threshold for the detector to value,
where value is a floating-point representation of the beta threshold. The
following example sets the beta threshold to 500.
Send:
set beta th a 500
Receive:
set beta th a 500 ok
Send:
set beta th b 500
Receive:
set beta th b 500 ok
bkg pm a/b
This command reports the current PM backgrounds. The following
example reports that the PM background is 4 µg/m
Send:
bkg pm a
bkg pm a 4.000E+00 m g/m3
Receive:
Send:
bkg pm b
bkg pm b 4.000E+00 m g/m3
Receive:
set bkg pm a/b value
This command is used to set PM backgrounds to user-defined values to
value, where value is a floating-point representation of the background
3
.