Table B-12. Analog Voltage Output Range Values - Thermo Scientific 5028i Instruction Manual

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
I/O Configuration
B-56
Model 5028i Instruction Manual
Send:
analog vin 1
Receive:
analog vin 1 75.325, 2.796 V
analog vout range channel
This command reports the analog voltage output channel range, where
channel is between 1 and 6 inclusive, according to
following example reports that analog voltage output channel 2 is set to
3 (0-10 V).
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range channel range
This command sets analog voltage output channel to the range, where
channel is between 1 and 6 inclusive, and range is set according to
B–12. The following example sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B–12. Analog Voltage Output Range Values
Range
1
2
3
4
0 [cannot be set to this, but may report]
dig in
This command reports the status of the digital inputs as a 4-digit
hexadecimal string with the most significant bit (MSB) being input 16.
Send:
dig in
Receive:
dig in 0xffff
din channel
This command reports the action assigned to the digital input
channel and the index number of the corresponding active state. The
following example reports input 1 to be assigned an index number 3
corresponding to action of "PM mode" with the active state being high.
Send:
din 1
Receive:
din 1 3 PM MODE high
Table
B–12. The
Output Range
0-1 V
0-100 mV
0-10 V
0-5 V
Undefined