Thermo Scientific 5028i Instruction Manual page 34

Continuous particulate monitor
Table of Contents

Advertisement

Introduction
Specifications
1-8
Model 5028i Instruction Manual
Two 3-foot heated sample tubes
Ambient T/RH Assembly
Consumables for average first year's operation (ambient)
iPort Software for Local or Remote Communication
Sample Heater
Two 3-foot insulated heater tube
Three Modes of Heater Control
Instrument Performance (16.67 l/min, 1s, stable conditions)
Measurement Range:
Minimum Detection Limit:
3
Resolution: 0.1 μg/m
3
Precision: ±3.0 μg/m
<80 μg/m
RMS Precision: PM2.5 < 5%, PM10 < 5%(24-hour)
Accuracy for Mass Measurement: ±5% using NIST-traceable mass foil set
Data Averaging and Output
Real-time Mass Conc Average: 20 minutes
Long-Term Averaging: 60 to 3,600 sec and 24-hour
Data Output Rate: every 1 second
Operating Range
The temperature of the sampled air may vary between -30 and 45 °C. The 5028i units
must be weather protected within the range of 4 °C to 50 °C.
An optional Complete Outdoor Enclosure provides complete weather protection.
Non-condensing; <95% RH inside 5028i
Sample Flow
Volumetric flow control system uses a calibrated subsonic precision orifice and the
measured ambient temperature and pressure to maintain constant volumetric flow rates.
Nominal Flow Rate: 16.67 L/min
Flow Precision ±2% of measured value
Flow Accuracy <5% of measured value
Data Storage
Internal data logging of user-specified variables; capacity of 190,000 records.
OFF – Heater is turned off in non condensing
environment
TEMP – fixed sample temperature
RH – Dynamic Heating of sample at set sample RH
threshold (e.g., 50% RH)
0 to 0.1, 1.0, 2.0, 3.0, 5.0, 10.0 mg/m
0 to 100, 1,000, 2,000, 3,000, 5,000, 10,000 μg/m
3
(1 hour) @ 2 s
<4 μg/m
3
<1 μg/m
(24-hour) @ 2 s
3
3
3
; 4-5 μg/m
>80 μg/m
(24-hour)
3
3
Thermo Fisher Scientific