Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-6
Model 5028i Instruction Manual
The Gesytec serial number defaults to zero. To set the Gesytec serial
number select Main Menu > Instrument Controls > Communication
Settings > Gesytec Serial No.
Example of response to DA command from an instrument with
Gesytec serial number set to 000. The Gesytec serial number is bold in
the example.
Gesytec Protocol with transmission of three concentrations (Instrument
ID is 1, Operation Status is 03, Error Status is 04):
Data Query String:
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>
Address First Concentration(E-format)=25.78
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681
000
0000000<SP><CR>
Example of response to DA command from an instrument with
Gesytec serial number set to 123. The Gesytec serial number is bold in
the example.
Gesytec Protocol with transmission of three concentrations (Instrument
ID is 1, Operation Status is 03, Error Status is 04):
Data Query String:
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>
Address First Concentration(E-format)=25.78
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681
123
0000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response, if not valid in a
particular mode, will report a value of 0.0.
<STX>DA<CR>
Address+2
<STX>DA<CR>
Address+2
000
0000000 <SP>002 <SP>
Address+1
Third Concentration=11.75
123
0000000 <SP>002 <SP>
Address+1
Third Concentration=11.75