Diagnostics
Thermo Fisher Scientific
Send:
set alarm vac pres max b 250.0
Receive:
set alarm vac pres max b 250.0 ok
amb rh
This command reports the ambient RH, in percent. The following
example reports that the ambient RH is 30%.
Send:
amb rh
Receive:
amb rh 30.0%
amb temp
This command reports the current ambient temperature, in degrees C.
The following example reports that the ambient temperature is 15 °C.
Send:
amb temp
Receive:
amb temp 15.0 degC
baro pres
This command reports the current barometric pressure, in mmHg. The
following example reports that the barometric pressure is 722 mmHg.
Send:
baro pres
Receive:
baro pres 722.0 mmHg
board temp
This command reports the current board temperature, in degrees C. The
following example reports that the board temperature is 15 °C.
Send:
board temp
Receive:
board temp 15.0 degC
crn a/b
This command reports the current radon concentration for the detector.
The following example reports that the radon concentration is 1.2 Bq/
3
m
.
Send:
crn a
Receive:
crn a 1.20 Bq/m3
Send:
crn b
Receive:
crn b 1.20 Bq/m3
det status a/b
This command reports the current mass for the detector. The sequence
is: β c, β , α . Each value is separated by a space.
Send:
det status a
Receive:
det status a 30000 25000 500
Send:
det status b
Receive:
det status b 30000 25000 500
C-Link Protocol Commands
Diagnostics
Model 5028i Instruction Manual
B-19