Operation
Alarms Menu

Min and Max Flow Limits

Concentration Alarms

3-68
Model 5028i Instruction Manual

Flow

The Flow screen displays the current flow and sets the minimum and
maximum alarm limits. Acceptable alarm limits range from 16.00
to 17.34 LPM. If the flow reading goes beyond the minimum or
maximum limit, an alarm is activated and the alarm (bell) icon appears
in the status bar on the Run screen and in the Main Menu.
Note Flow B settings please refer to Flow as following description.
In the Main Menu, choose Alarms > Flow Alarms > Flow or Flow B.
l
The Minimum Flow alarm limit screen is used to change the minimum
flow alarm limit. The minimum and maximum flow screens function
the same way.
In the Main Menu, choose Alarms > Flow Alarms > Flow or Flow B
l
> Min or Max.
The Concentration Alarms submenu allows the user to view and set the
24-hour average PM and instant PM concentration alarm parameters.
In the Main Menu, choose Alarms > Conc Alarms.
l
FLOW:
ACTUAL
>MIN
MAX
RANGE
AVG
DIAGS ALARM
FLOW:
ACT MIN:
SET MIN:
RANGE
AVG
DIAGS ALARM
CONC ALARMS:
>AVG PM
AVG PMb
INST PM
INST PMb
RANGE
AVG
DIAGS ALARM
0.00 LPM
16.17 LPM
17.17 LPM
16.00 LPM
16.17 LPM?
INC/DEC
SAVE VALUE
OK
OK
OK
OK