Thermo Scientific 5028i Instruction Manual page 137

Continuous particulate monitor
Table of Contents

Advertisement

Analog Input Calibrate Zero
Analog Input Calibrate Full-Scale
Thermo Fisher Scientific
The Analog Input Calibrate Zero screen allows the user to calibrate the
zero state of the selected analog input.
In the Main Menu, choose Service > Analog Input Cal > select
l
Channel > Calibrate Zero.
The Analog Input Calibrate Full-Scale screen allows the user to calibrate
the full-scale state of the selected analog input.
In the Main Menu, choose Service > Analog Input Cal > select
l
Channel > Calibrate Full Scale. (Hook up a voltage source of 10 V
to the analog input channel.)
Language
The Language menu allows the user to choose languages include
Chinese and English.
In the Main Menu, choose Service>Language.
l
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
RANGE
AVG
DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE
AVG
DIAGS ALARM
ANALOG INPUT CAL:
DISCONNECT SELECTED INPUT!
SELECTED INPUT:
CURRENTLY:
CALIBRATE INPUT TO ZERO
RANGE
AVG
DIAGS ALARM
ANALOG INPUT CAL:
PROVIDE VOLTAGE TO INPUT!
SELECTED INPUT:
CURRENTLY:
SET TO:
CALIBRATE TO VALUE
RANGE
AVG
DIAGS ALARM
Operation
Service Menu
ZERO
INPUT 1
0.04 V
SPAN
INPUT 1
9.84 V
10.00
V
Model 5028i Instruction Manual
3-83