Thermo Scientific 5028i Instruction Manual page 275

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
flow pres offset
This command reports the flow pressure offset for the detector at the
time of calibration. The following example reports that the flow pressure
offset is 3.0.
Send:
flow pres offset
Receive:
flow pres offset 3.0
flow span a/b
This command reports the current flow span at the time of calibration.
The following example reports that the flow at calibration is 1.00.
Send:
flow span a
Receive:
flow span a 1.000
Send:
flow span b
Receive:
flow span b 1.000
set flow span a/b value
This command sets the current flow span to the value, where value is a
floating-point representation of the flow span specified at the time of
calibration. The following example sets the flow span to 1.5.
Send:
set flow span a 1.5
Receive:
set flow span a 1.5 ok
Send:
set flow span b 1.5
Receive:
set flow span b 1.5 ok
flow temp offset a/b
This command reports offset for flow temperature, in degrees C. The
following example reports that the flow temperature offset is 5 °C.
Send:
flow temp offset a
Receive:
flow temp offset a 5.0 degC
Send:
flow temp offset b
Receive:
flow temp offset b 5.0 degC
set flow temp offset a/b value
This command sets the offset value for flow temperature, where value is
a floating-point number representing flow temperature offset in degrees
C. The following example sets the flow temperature offset to 7 °C.
Send:
set flow temp offset a 7
Receive:
set flow temp offset a 7 ok
Send:
set flow temp offset b 7
Receive:
set flow temp offset b 7 ok
C-Link Protocol Commands
Calibration
Model 5028i Instruction Manual
B-35