Thermo Scientific 5028i Instruction Manual page 287

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
Level
Contrast Level
16
80%
17
85%
18
90%
19
95%
20
100%
date
This command reports the current date. The following example reports
the date as April 1, 2009.
Send:
date
Receive:
date 04-01-09
set date mm-dd-yy
mm = month
dd = day
yy = year
This command sets the date of the analyzer's internal clock. The
following example sets the date to May 1, 2009.
Send:
set date 05-01-09
Receive:
set date 05-01-09 ok
set default params
This command sets all the parameters to their default values. This does
not affect the factory-calibrated parameters.
Send:
set default params
Receive:
set default params ok
det
This command reports the status of the detector board (sharp/beta/map/
none). The following example reports that the detector is beta.
Send:
det
Receive:
det beta
pump
This command reports the current status of pump as on or off. The
following example reports that the pump is set to off.
Send:
pump
Receive:
pump off
C-Link Protocol Commands
Hardware Configuration
Model 5028i Instruction Manual
B-47