Thermo Scientific ALPHA COND 550 Manual

Conductivity monitor ms/µs/ºc/ºf meter
Hide thumbs Also See for ALPHA COND 550:

Advertisement

Quick Links

Alpha CON 550
Conductivity Monitor
mS/µS/ºC/ºF meter

Advertisement

Table of Contents
loading

Summary of Contents for Thermo Scientific ALPHA COND 550

  • Page 1 Alpha CON 550 Conductivity Monitor mS/µS/ºC/ºF meter...
  • Page 3 Preface This manual serves to explain the use of the Alpha-CON 550 Industrial Conductivity Monitor. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide.
  • Page 5: Table Of Contents

    TABLE OF CONTENTS 1. Introduction ......................1 Before You Begin ........................1 Intended Use .......................... 1 Safety Instructions........................2 Taking Out of Service / Correct Disposal of the Unit ............. 2 2. Getting Started ......................3 Description of Instrument ....................... 3 Measurement System ......................4 Connecting Peripherals......................
  • Page 7: Introduction

    Instruction Manual Alpha-CON 550 1. Introduction 1.1 Before You Begin lpha We thank you for purchasing the A CON 550 monitor. The construction of the CON 550 monitor employs leading edge technology and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment.
  • Page 8: Safety Instructions

    Instruction Manual Alpha-CON 550 1.3 Safety Instructions - The Alpha CON 550 monitor should be installed and operated only by personnel familiar with the instrument and who are qualified for such work. - A defective CON monitor must neither be installed nor put into service.
  • Page 9: Getting Started

    Instruction Manual Alpha-CON 550 2. Getting Started 2.1 Description of Instrument The Alpha CON 550 monitor is used for measuring conductivity and temperature values. The monitor can be used for applications such as water treatment and controlling, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
  • Page 10: Measurement System

    Instruction Manual Alpha-CON 550 2.2 Measurement System A typical measurement system consists of: • A conductivity process monitor • A conductivity sensor with integrated or separate Pt100/Pt1000 temperature sensor • An appropriate measurement cable • An immersion, flow or process assembly Alpha CON 550 Monitor lpha pH560 pH Monitor...
  • Page 11: Connecting Peripherals

    Instruction Manual Alpha-CON 550 2.3 Connecting Peripherals 2.3.1 Connection Terminals Remove Back Cover: Remove the screws from the four corners at the back of the conductivity monitor. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below.
  • Page 12: Switching Between Pt100 & Pt1000 Temperature Sensors

    Instruction Manual Alpha-CON 550 2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors The conductivity monitor supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below.
  • Page 13: Connecting Conductivity Probe

    Instruction Manual Alpha-CON 550 2.3.3 Connecting Conductivity Probe Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity monitor. 2-Cell Conductivity Probe: 1. Short Pin 1 & 2 of J10 connector using a jumper wire. 2. Connect CON low input wire (white) to Pin 2 of J10 connector. 3.
  • Page 14: Installation

    Instruction Manual Alpha-CON 550 2.4 Installation 2.4.1 Mechanical Dimensions lpha CON550 Conductivity Monitor 2.4.2 Wall Mount Pierce through holes at both sides Cover the catch slots at both sides with overlays...
  • Page 15: Panel Mount

    Instruction Manual Alpha-CON 550 2.4.3 Panel Mount Prepare panel cut-out of 92.0 mm X 92.0 mm Remove back cover of conductivity monitor and slide it through panel cut-out Gasket Panel Panel Insert threaded rods through catch until conductivity monitor is held against panel Attach catch to both sides of conductivity monitor...
  • Page 16: Display & Keypad

    Instruction Manual Alpha-CON 550 2.5 Display & Keypad 2.5.1 Display Overview The Liquid Crystal Display (LCD) of αlpha CON 550 conductivity monitor has two alpha-numerical displays (Upper and a Lower). • Upper display: Measured conductivity value is displayed when the monitor is in normal operation (measurement) mode.
  • Page 17: Key Functions

    Instruction Manual Alpha-CON 550 Units of Measurement Indicators Conductivity in Milli Siemens Conductivity in Micro Siemens µS Line Resistance in Ohms (Refer Section 5.5) Temperature Coefficient (Refer Section 5.2) Temperature in Celsius ºC (Refer Section 5.3) Temperature in Fahrenheit ºF (Refer Section 5.3) 2.5.2 Key Functions lpha CON550...
  • Page 18: Operation

    Instruction Manual Alpha-CON 550 3. Operation 3.1 Measurement mode When the conductivity monitor is powered on, the display shows all the LCD segments briefly, and then automatically enters into the measurement mode. MEAS µS READY °C 2 5 . 0 The mode indicator ‘MEAS’...
  • Page 19: Menu Overview

    Instruction Manual Alpha-CON 550 3.2 Menu Overview MEAS µS READY °C SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD SETUP HOLD...
  • Page 20: Calibration Mode

    Instruction Manual Alpha-CON 550 4. Calibration Mode 4.1 About Calibration Calibration should be carried out each time a new electrode is attached to the monitor or when you suspect that the monitor/electrode is out of calibration. The monitor allows you to perform temperature calibration and conductivity calibration.
  • Page 21: Calibration

    Instruction Manual Alpha-CON 550 4.2 Calibration The monitor requires one-point calibration. Conductivity calibration is always carried out in the specific measuring range selected for the monitor. Make sure, the monitor is set to desired measuring range before you enter calibration mode (Refer Section 5.4 for selecting measuring ranges). Select a suitable standard conductivity solution so that its conductivity value is within 10% to 100% of the full scale of the selected measuring range of the monitor.
  • Page 22 Instruction Manual Alpha-CON 550 NOTES: • The acceptable calibration window is ±40% of the displayed (default) value. For example, if the display is 1000 µS, the values to which it can be adjusted is 600 to 1400 µS. • If the conductivity reading of the solution is not within 10% to 100% of the full scale of the selected measuring range, calibration error occurs;...
  • Page 23: Setup Mode

    Instruction Manual Alpha-CON 550 5. Setup Mode 5.1 Enter Setup mode The setup mode allows you to customize the settings of the conductivity monitor to suite your requirements. While in measurement mode, press the ENT key to access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup (rAng – Range settings).
  • Page 24: Temperature Coefficient Settings

    Instruction Manual Alpha-CON 550 5.2 Temperature Coefficient Settings Setting the temperature coefficient to match your process/calibration liquids allows monitor to achieve conductivity readings with higher accuracy. SETUP SETUP SETUP SETUP HOLD HOLD HOLD HOLD SETUP HOLD From measurement mode press ENT key to enter setup mode as described in section 5.1.
  • Page 25: Temperature Settings

    Instruction Manual Alpha-CON 550 5.3 Temperature Settings Automatic Temperature Compensation (ATC): The conductivity value of a sample is affected by temperature. Use ATC feature of the monitor to compensate for conductivity changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the monitor if an integrated conductivity/Temperature probe is not available.
  • Page 26 Instruction Manual Alpha-CON 550 If ATC enabled (ATC On): Selecting temperature offset: The upper display shows the last configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner.
  • Page 27: Measuring Range Settings

    Instruction Manual Alpha-CON 550 5.4 Measuring Range Settings The conductivity monitor has 5 different measuring ranges. You need to select a range which is the most suitable for conductivity reading of your applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm). The measuring range you select here will affect the way the monitor behaves in the measurement mode and calibration mode.
  • Page 28: Configuration Settings

    Instruction Manual Alpha-CON 550 5.5 Configuration Settings Configuration settings let you select between 2-cell & 4-cell and reset the monitor to factory defaults. SETUP SETUP SETUP SETUP HOLD HOLD HOLD HOLD This screen appears only when 2-cell is selected and measuring range is set to 0.0 to 200.0 mS/cm SETUP SETUP...
  • Page 29 Instruction Manual Alpha-CON 550 measuring range is selected to ‘0.0 to 200.0 mS/cm’. Press ▲ or ▼ key to set the line resistance of the sensor cable connected to the monitor. Press ENT key to confirm your selection. Reset to factory defaults: The display shows ‘nO dEF’. Press ▲...
  • Page 30: Viewing Electrode Properties

    Instruction Manual Alpha-CON 550 5.6 Viewing Electrode Properties Each time you calibrate the conductivity sensor, the monitor re-calculates the calibration factor of the sensor and shown in the LCD at the end of calibration. The setup mode allows you to view the last saved calibration factor and cell constant at any time.
  • Page 31: Technical Specifications

    Instruction Manual Alpha-CON 550 6. Technical Specifications General Specification Conductivity Measuring Range: Resolution: 0.00 - 20.00 µS/cm 0.01 µS/cm 0.0 - 200.0 µS/cm 0.1 µS/cm 0- 2000 µS/cm 1 µS/cm 0.00 - 20.00 mS/cm 0.01 mS/cm 0.0 - 200.0 mS/cm 0.1 mS/cm Relative accuracy ±...
  • Page 32 Instruction Manual Alpha-CON 550 Calibration Default Settings (CAL dEF) Conductivity Calibration factor Reset to 1.000 Factory Default Settings (FCt dEF) Conductivity Calibration factor Reset to 1.000 Cell constant 1.00 Cell type 2-cell Measuring range 0 to 2000 µS/cm Line resistance 0.000 Ω...
  • Page 33: List Of Accessories

    Instruction Manual Alpha-CON 550 7. List of Accessories Monitor Replacement and Accessories Item Description Order Code Alpha CON 550 Monitor EC-PHCP0550 Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 0.1 EC-CONSEN89X Conductivity cell, Epoxy body, Graphite sensor, w/3-wire Pt100, k = 1.0 EC-CONSEN88X Conductivity 2 Cell type probe, 0.1 - 200μS;...
  • Page 34: Troubleshooting

    Instruction Manual Alpha-CON 550 8. Troubleshooting Cause Problem Solution Power on, but no display a) Loose connections a) Ensure cables make good contact b) Incorrect output voltage of the power adaptor b) Use an power adaptor with specified output voltage c) Cables not in correct polarity (+ and –...
  • Page 35: General Information

    Dept. has to be informed in advance. Items must be carefully packed to prevent damage during shipment, and insured against possible damage or loss. Eutech Instruments will not be responsible for any damage resulting from careless or insufficient packing. Warning: Shipping damage as a result of inadequate packaging is the user's/distributor’s responsibility.
  • Page 36: Appendices

    Instruction Manual Alpha-CON 550 10. Appendices 10.1 Appendix 2 Conductivity of various aqueous solutions At Temperature 25 C /77 Solution Conductivity Resistivity Ultra-pure Water 0.055 µS/cm 18.18 MΩ-cm Power Plant Boiler Water 0.05 - 1 µS/cm 1 - 18 MΩ-cm Distilled Water 0.5 µS/cm 2 MΩ-cm...
  • Page 37: Appendix 4 - Abbreviations Used In Lcd

    Instruction Manual Alpha-CON 550 10.2 Appendix 4 - Abbreviations Used in LCD Abbreviation Description Automatic Temperature Compensation Calibration mode C.ºC Calibration temperature CdAt Electrode properties Cell CnFg Configuration C.tC Temperature coefficient for calibration liquid Default values Factory (defaults) HOLD Relay functions are hold Cell constant L.Ad Linear resistance...
  • Page 38 NOTES...
  • Page 40 Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int’l Fax: 978-232-6031 Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949...

This manual is also suitable for:

Alpha con 550

Table of Contents