Preface This manual serves to explain the use of the Alpha-CON 550 Industrial Conductivity Monitor. The manual functions in two ways, firstly as a step by step guide to help the user operate the instrument. Secondly, it serves as a handy reference guide.
TABLE OF CONTENTS 1. Introduction ......................1 Before You Begin ........................1 Intended Use .......................... 1 Safety Instructions........................2 Taking Out of Service / Correct Disposal of the Unit ............. 2 2. Getting Started ......................3 Description of Instrument ....................... 3 Measurement System ......................4 Connecting Peripherals......................
Instruction Manual Alpha-CON 550 1. Introduction 1.1 Before You Begin lpha We thank you for purchasing the A CON 550 monitor. The construction of the CON 550 monitor employs leading edge technology and complies with safety regulations currently in force. Notwithstanding this, improper use could lead to hazards for the user or a third-party, and/or adverse effects on the plant or other equipment.
Instruction Manual Alpha-CON 550 1.3 Safety Instructions - The Alpha CON 550 monitor should be installed and operated only by personnel familiar with the instrument and who are qualified for such work. - A defective CON monitor must neither be installed nor put into service.
Instruction Manual Alpha-CON 550 2. Getting Started 2.1 Description of Instrument The Alpha CON 550 monitor is used for measuring conductivity and temperature values. The monitor can be used for applications such as water treatment and controlling, galvanic-decontamination, chemical processing, food processing, clean or wastewater control and neutralization processes.
Instruction Manual Alpha-CON 550 2.2 Measurement System A typical measurement system consists of: • A conductivity process monitor • A conductivity sensor with integrated or separate Pt100/Pt1000 temperature sensor • An appropriate measurement cable • An immersion, flow or process assembly Alpha CON 550 Monitor lpha pH560 pH Monitor...
Instruction Manual Alpha-CON 550 2.3 Connecting Peripherals 2.3.1 Connection Terminals Remove Back Cover: Remove the screws from the four corners at the back of the conductivity monitor. Remove the back cover. The connectors are exposed on the back PCBA as shown in the Figure 1 below.
Instruction Manual Alpha-CON 550 2.3.2 Switching Between Pt100 & Pt1000 Temperature Sensors The conductivity monitor supports both Pt100 & Pt1000 (2-wire or 3-wire) temperature sensors. The default factory setting is Pt100. If you need to use Pt1000 temperature sensor, you have to change the jumper setting (J7) as described below.
Instruction Manual Alpha-CON 550 2.3.3 Connecting Conductivity Probe Any industrial conductivity probe (2-cell or 4-cell) can be connected to the conductivity monitor. 2-Cell Conductivity Probe: 1. Short Pin 1 & 2 of J10 connector using a jumper wire. 2. Connect CON low input wire (white) to Pin 2 of J10 connector. 3.
Instruction Manual Alpha-CON 550 2.4 Installation 2.4.1 Mechanical Dimensions lpha CON550 Conductivity Monitor 2.4.2 Wall Mount Pierce through holes at both sides Cover the catch slots at both sides with overlays...
Instruction Manual Alpha-CON 550 2.4.3 Panel Mount Prepare panel cut-out of 92.0 mm X 92.0 mm Remove back cover of conductivity monitor and slide it through panel cut-out Gasket Panel Panel Insert threaded rods through catch until conductivity monitor is held against panel Attach catch to both sides of conductivity monitor...
Instruction Manual Alpha-CON 550 2.5 Display & Keypad 2.5.1 Display Overview The Liquid Crystal Display (LCD) of αlpha CON 550 conductivity monitor has two alpha-numerical displays (Upper and a Lower). • Upper display: Measured conductivity value is displayed when the monitor is in normal operation (measurement) mode.
Instruction Manual Alpha-CON 550 Units of Measurement Indicators Conductivity in Milli Siemens Conductivity in Micro Siemens µS Line Resistance in Ohms (Refer Section 5.5) Temperature Coefficient (Refer Section 5.2) Temperature in Celsius ºC (Refer Section 5.3) Temperature in Fahrenheit ºF (Refer Section 5.3) 2.5.2 Key Functions lpha CON550...
Instruction Manual Alpha-CON 550 3. Operation 3.1 Measurement mode When the conductivity monitor is powered on, the display shows all the LCD segments briefly, and then automatically enters into the measurement mode. MEAS µS READY °C 2 5 . 0 The mode indicator ‘MEAS’...
Instruction Manual Alpha-CON 550 4. Calibration Mode 4.1 About Calibration Calibration should be carried out each time a new electrode is attached to the monitor or when you suspect that the monitor/electrode is out of calibration. The monitor allows you to perform temperature calibration and conductivity calibration.
Instruction Manual Alpha-CON 550 4.2 Calibration The monitor requires one-point calibration. Conductivity calibration is always carried out in the specific measuring range selected for the monitor. Make sure, the monitor is set to desired measuring range before you enter calibration mode (Refer Section 5.4 for selecting measuring ranges). Select a suitable standard conductivity solution so that its conductivity value is within 10% to 100% of the full scale of the selected measuring range of the monitor.
Page 22
Instruction Manual Alpha-CON 550 NOTES: • The acceptable calibration window is ±40% of the displayed (default) value. For example, if the display is 1000 µS, the values to which it can be adjusted is 600 to 1400 µS. • If the conductivity reading of the solution is not within 10% to 100% of the full scale of the selected measuring range, calibration error occurs;...
Instruction Manual Alpha-CON 550 5. Setup Mode 5.1 Enter Setup mode The setup mode allows you to customize the settings of the conductivity monitor to suite your requirements. While in measurement mode, press the ENT key to access setup mode. LCD shows ‘SETUP’ mode indicator and the first page of setup (rAng – Range settings).
Instruction Manual Alpha-CON 550 5.2 Temperature Coefficient Settings Setting the temperature coefficient to match your process/calibration liquids allows monitor to achieve conductivity readings with higher accuracy. SETUP SETUP SETUP SETUP HOLD HOLD HOLD HOLD SETUP HOLD From measurement mode press ENT key to enter setup mode as described in section 5.1.
Instruction Manual Alpha-CON 550 5.3 Temperature Settings Automatic Temperature Compensation (ATC): The conductivity value of a sample is affected by temperature. Use ATC feature of the monitor to compensate for conductivity changes when the temperature of the sample or process liquid fluctuates. Connect a separate temperature probe (Pt100/pt1000) to the monitor if an integrated conductivity/Temperature probe is not available.
Page 26
Instruction Manual Alpha-CON 550 If ATC enabled (ATC On): Selecting temperature offset: The upper display shows the last configured temperature offset value (if any), otherwise the default is zero. The lower display shows currently measured temperature reading (including last configured offset value). LCD shows ‘ATC’ annunciator in lower-right corner.
Instruction Manual Alpha-CON 550 5.4 Measuring Range Settings The conductivity monitor has 5 different measuring ranges. You need to select a range which is the most suitable for conductivity reading of your applications/samples. The factory default is 2000 µS (0 to 2000 µS/cm). The measuring range you select here will affect the way the monitor behaves in the measurement mode and calibration mode.
Instruction Manual Alpha-CON 550 5.5 Configuration Settings Configuration settings let you select between 2-cell & 4-cell and reset the monitor to factory defaults. SETUP SETUP SETUP SETUP HOLD HOLD HOLD HOLD This screen appears only when 2-cell is selected and measuring range is set to 0.0 to 200.0 mS/cm SETUP SETUP...
Page 29
Instruction Manual Alpha-CON 550 measuring range is selected to ‘0.0 to 200.0 mS/cm’. Press ▲ or ▼ key to set the line resistance of the sensor cable connected to the monitor. Press ENT key to confirm your selection. Reset to factory defaults: The display shows ‘nO dEF’. Press ▲...
Instruction Manual Alpha-CON 550 5.6 Viewing Electrode Properties Each time you calibrate the conductivity sensor, the monitor re-calculates the calibration factor of the sensor and shown in the LCD at the end of calibration. The setup mode allows you to view the last saved calibration factor and cell constant at any time.
Instruction Manual Alpha-CON 550 8. Troubleshooting Cause Problem Solution Power on, but no display a) Loose connections a) Ensure cables make good contact b) Incorrect output voltage of the power adaptor b) Use an power adaptor with specified output voltage c) Cables not in correct polarity (+ and –...
Dept. has to be informed in advance. Items must be carefully packed to prevent damage during shipment, and insured against possible damage or loss. Eutech Instruments will not be responsible for any damage resulting from careless or insufficient packing. Warning: Shipping damage as a result of inadequate packaging is the user's/distributor’s responsibility.
Instruction Manual Alpha-CON 550 10.2 Appendix 4 - Abbreviations Used in LCD Abbreviation Description Automatic Temperature Compensation Calibration mode C.ºC Calibration temperature CdAt Electrode properties Cell CnFg Configuration C.tC Temperature coefficient for calibration liquid Default values Factory (defaults) HOLD Relay functions are hold Cell constant L.Ad Linear resistance...