Thermo Scientific 5028i Instruction Manual page 114

Continuous particulate monitor
Table of Contents

Advertisement

Operation
Alarms Menu
Max Filter Tape Counter
Detector Alarms
Min and Max Alpha Alarms
3-60
Model 5028i Instruction Manual
In the Main Menu, choose Alarms > Instrument Alarms > Filter
l
Tape Counter.
FILTER TAPE COUNTER:
ACTUAL
>MAX
RANGE
The Maximum Filter Tape Counter screen is used to change the
maximum filter tape counter alarm limit (1200 = 90% usage, 10%
remaining).
In the Main Menu, choose Alarms > Instrument Alarms > Filter
l
Tape Counter > Max.
FILTER TAPE COUNTER:
ACT MAX:
SET MAX:
CHANGE VALUE
RANGE
The Detector Alarms submenu allows the user to view a set of alarms
for detector boards. Board status indicates that the power supplies are
working and connections are successful. There are no setting screens for
this alarm.
In the Main Menu, choose Alarms > Detector Alarms.
l
DETECTOR ALARMS:
BOARD STATUS
BOARD STATUS
>ALPHA
ALPHA B
BETA
BETA B
RANGE
The Alpha Alarms screen allows the user to view the minimum and
maximum alarm parameters. Acceptable alarm limits range from 0 to
100. If the alpha count reading goes beyond the maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on
the Run screen and in the Main Menu
Note Alpha B settings please refer to Alpha as following description.
AVG
DIAGS ALARM
450
xxxx
1300
MOVE CURSOR
SAVE
AVG
DIAGS ALARM
B
AVG
DIAGS ALARM
0
1300
0
?
OK
OK
OK
OK
OK