Model 5028i Instruction Manual
One limiting factor when siting the Model 5028i is placement within
a standard 19-inch rack mounting. Future planning should be made
to reserve the topmost rack mounting position available due to the
vertical positioning of the sample tube directly from the roof into the
top of the Model 5028i. In addition, modification to the rack cabinet to
accommodate the vertical tubing connection also needs to be considered
prior to installation. For more information, see Chapter 9
Equipment".
Review the following installation steps prior to field installation.
Once all height requirements are established, the following steps are
usually taken to complete the installation.
1. Open the front door of 5028i and inspect for proper filter tape
installation.
2. Place 5028i unit in place of intended operation.
3. Attach the insulated heater tube assembly to the 5/8-inch OD tubes
on top of the 5028i
To Rear Panel 5028i "CHA
Fasten Nut
HEATER" or
"CHB HEATER"
Figure 2–5. Heater Assembly
4. Attach a second tubing connector to the top of the heater tube.
(Figure
2–5).
Tubing Connector
"Optional