Analog Output Configuration
(Select Channel)
Allow Over/Under Range
Thermo Fisher Scientific
CHOOSE ACTION:
>NONE
MODE A
MODE B
MODE AB
AOUTS TO ZERO
AOUTS TO FS
FT COUNT
RANGE
FT CNTRL
PUMP CNTRL
TEMP CNTRL
PRES CNTRL
HT CNTRL
RH/TEMP CNTRL
HT CNTRL B
RH/TEMP CNTRL
EXT ALARM 1
EXT ALARM 2
EXT ALARM 3
FT COUNT B
FTCNTRL B
The Analog Output Configuration menu displays a list of the analog
output channels available for configuration. Channel choices include all
voltage channels, all current channels, individual voltage channels 1–6,
and individual current channels 1–6 (if the I/O expansion board option
is installed).
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Output Config.
OUTPUT CHANNELS:
>ALLOW OVER/UNDER RANGE
ALL VOLTAGE CHANNELS
VOLTAGE CHANNEL 1
VOLTAGE CHANNEL 2
VOLTAGE CHANNEL 3
RANGE
The Allow Over/Under Range screen is used to select whether or not
the analog outputs are allowed to exceed the maximum selected value of
100 mV, 1 V, 5 V, 10 V, or 20 mA or the minimum selected value of 0 V,
0 mA, or 4 mA. By default, this parameter is set to on and 5% over and
under range is allowed for all analog output channels.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Output Config > Allow Over/Under
Range.
AVG
DIAGS ALARM
B
AVG
DIAGS ALARM
Model 5028i Instruction Manual
Operation
Instrument Controls Menu
<--
3-43