Ambient Temperature
Min and Max Ambient
Temperature Limits
Flow Temperature
Thermo Fisher Scientific
In the Main Menu, choose Alarms > RH/Temperature Alarms >
l
Sample RH or Sample RH B > Min or Max.
SAMPLE RH:
ACT MIN:
SET MIN:
RANGE
The Ambient Temperature screen displays the current ambient
temperature value and sets the minimum and maximum alarm limits.
Acceptable alarm limits range from -30 °C to 60 °C. If the ambient
temperature reading goes beyond the minimum or maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on
the Run screen and in the Main Menu.
In the Main Menu, choose Alarms > RH/Termperature Alarms >
l
Ambient Temp.
AMBIENT TEMP:
ACTUAL
>MIN
MAX
RANGE
The Minimum Ambient Temperature alarm limit screen is used to
change the minimum ambient temperature alarm limit. The minimum
and maximum ambient temperature screens function the same way.
In the Main Menu, choose Alarms > RH/Temperature Alarms >
l
Ambient Temp > Min or Max.
AMBIENT TEMPERATURE:
ACT MIN:
SET MIN:
RANGE
The Flow Temperature screen displays the current ambient temperature
value and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 5 to 60 °C. If the flow temperature reading goes
beyond the minimum or maximum limit, an alarm is activated and the
alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu
5.0 %
35 %?
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
-30.0
-30.0
60.0
AVG
DIAGS ALARM
-30.0
-20.0
INC/DEC
SAVE VALUE
AVG
DIAGS ALARM
Model 5028i Instruction Manual
Operation
Alarms Menu
o
C
o
C
o
C
o
C
o
C?
3-63