Range Menu; Concentration Units - Thermo Scientific 5028i Instruction Manual

Continuous particulate monitor
Table of Contents

Advertisement

Range Menu

Concentration Units

Thermo Fisher Scientific
PM
PMb
x
MAIN MENU:
>RANGE
INTEGRATION
24HR AVERAGE
CALIBRATION FACTORS
INSTRUMENT CONTROLS
DIAGNOSTICS
ALARMS
RANGE
SERVICE
PASSWOR
The Range menu allows the operator to select the concentration units,
PM/PM
standard ranges, and to set the custom ranges.
b
Note PM
Range settings please refer to PM Range as following
b
description.
s
In the Main Menu, choose Range
l
RANGE:
>CONC UNITS
PM RANGE
PM
b
SET CUSTOM
RANGE
The Concentration Units screen defines how the particles concentration
readings are expressed. Concentration units of micrograms per cubic
3
meter (µg/m
) and milligrams per cubic meter (mg/m
When switching the selected units, the instrument's analog ranges all
default to the highest range. For example, when switching from mg/m
3
to µg/m
, the ranges default to 10000 µg/m
change units, you should also check the range settings.
Note If the units change, the instrument should be re-calibrated. A
display warning will appear.
In the Main Menu, choose Range > Conc Units.
l
CONCENTRATION
15.0
15.0 µg/m3
12:34
TIME
AVG
DIAGS ALARM
RANGE
RANGES
AVG
DIAGS ALARM
3
. Therefore, whenever you
s
Operation
Range Menu
µg/m3
3
x
μg/m3
10000
10000
3
) are available.
Model 5028i Instruction Manual
3
3-9