Data Content

Data Content

Shutdown

Important Tips

2-18
Model 5028i Instruction Manual
5-minute logging with 70% memory allocated. If there are
any changes, be sure to Commit Content.
b. From the Main Menu, choose Service. Under Service:
Note If Service Mode is not displayed, refer to
Mode" on page 7-4, then return to the beginning of this step.
i. Check/Calibrate Ambient Temperature, Sample Temperature,
and Flow Temperature
ii. Check/Calibrate Ambient RH
iii. Check Calibrate Barometric Pressure Span
iv. Check/Calibrate Flow Calibration
v. Perform a Auto Detector Calibration (30 minutes)
vi. Perform a Mass Foil Calibration (factory calibrated)
vii. Mass Coefficients are set at the factory.
Please use iPort to access data download from the instrument.
Use the following procedure when shutting down the instrument.
1. Turn the power OFF.
2. Cover instrument ports when not in use.
DO NOT
Do not pressurize the monitor.
l
Do not allow liquids to enter the sample port.
l
Do not expose the instrument to excessive vibration or magnetic
l
interference.
Do not remove/adjust the beta attenuation bench of the 5028i.
l
Do not remove the C-14 radioactive source at any time.
l
DO
Operate the instrument in a climate-controlled environment
l
(4–50 °C).
Set DHCP to ON when changing the IP address.
l
"Accessing the Service
s