C-Link Protocol Commands
Record Layout Definition
Record Layout
Definition
Format Specifier for ASCII
Responses
Format Specifier for Binary
Responses
B-60
Model 5028i Instruction Manual
Note If the command is sent without an appended relay number, then
all the relays are assigned the set logic of normally open/closed.
Send:
set relay open 1
Receive:
set relay open 1 ok
The erec, lrec, and srec layouts contain the following:
A format specifier for parsing ASCII responses
l
A format specifier for parsing binary responses
l
In addition to these, the erec layout contains:
A format specifier for producing the front panel display screens
l
Values are read using either the ASCII or binary format specifiers and
converted to uniform internal representations (32-bit floats or 32-bit
integers). These values are converted into text for display on the screen
using the format specifier for the front panel display. Normally, the
specifier used to parse a particular datum from the input stream will
be strongly related to the specifier used to display it (such as, all of the
floating point inputs will be displayed with an 'f ' output specifier, and
all of the integer inputs will be displayed with a 'd' specifier).
The first line of the layout response is the scanf-like parameter list for
parsing the fields from an ASCII erec response. Parameters are separated
by spaces and the line is terminated by a "\n" (the normal line separator
character). Valid fields are:
%s - parse a string
%d - parse a decimal number
%ld - parse a long (32-bit) decimal number
%f - parse a floating point number
%x - parse a hexadecimal number
%lx - parse a long (32-bit) hex number
%* - ignore the field
Note Signed versus unsigned for the integer values does not matter; it is
handled automatically.
The second line of the layout response is the binary parameter list
for parsing the fields from a binary response. Parameters MUST be
separated by spaces, and the line is terminated by a '\n'. Valid fields are:
t - parse a time specifier (2 bytes)
D - parse a date specifier (3 bytes)
i - ignore one 8-bit character (1 byte)
e - parse a 24-bit floating point number (3 bytes: n/x)
E - parse a 24-bit floating point number (3 bytes: N/x)
s
s
Thermo Fisher Scientific