Thermo Scientific 5028i Instruction Manual page 270

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
Calibration
Calibration
B-30
Model 5028i Instruction Manual
alpha eff a/b
This command reports the current value of alpha efficiency for the
detector. The following example reports that the alpha efficiency is 0.12.
Send:
alpha eff a
Receive:
alpha eff a 0.120
Send:
alpha eff b
Receive:
alpha eff b 0.120
set alpha eff a/b value
This command sets the alpha efficiency for the detector to value, where
value is a floating-point number representing the alpha efficiency for the
detector. The following example sets the alpha efficiency to.0.15.
Send:
set alpha eff a 0.15
Receive:
set alpha eff a 0.150 ok
Send:
set alpha eff b 0.15
Receive:
set alpha eff b 0.150 ok
alpha th a/b
This command reports the current value of alpha threshold for the
detector. The following example reports that the alpha threshold is
1220.
Send:
alpha th a
Receive:
alpha th a 1220
Send:
alpha th b
Receive:
alpha th b 1220
set alpha th a/b value
This command sets the alpha threshold for the detector to value, where
value is a floating-point number representing the alpha threshold for the
detector. The following example sets the alpha threshold to 1250.
Send:
set alpha th a 1250
Receive:
set alpha th a 1250 ok
Send:
set alpha th b 1250
Receive:
set alpha th b 1250 ok
amb rh offset
This command reports the current ambient RH offset, in percent. The
following example reports that the ambient RH offset is 0.6%.
Send:
amb rh offset
Receive:
amb rh offset 0.6%