Thermo Scientific 5028i Instruction Manual page 288

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
Communications Configuration
Communications
Configuration
B-48
Model 5028i Instruction Manual
set pump onoff
onoff = | on | off |
This command sets the pump on or off. The following example sets the
pump to on.
Send:
set pump on
Receive:
set pump on ok
save
set save params
These commands store all current parameters in FLASH memory. It
is important that each time instrument parameters are changed, that
this command be sent. If changes are not saved, they will be lost in the
event of a power failure. The following example saves the parameters to
FLASH memory.
Send:
set save params
Receive:
set save params ok
time
This command reports the current time (24-hour format). The following
example reports that the internal time is 2:15:30 pm.
Send:
time
Receive:
time 14:15:30
set time hh:mm:ss
hh = hours
mm = minutes
ss = seconds
This command sets the internal clock (24-hour format). The following
example sets the internal time to 2:15 pm.
Note If seconds are omitted, the seconds default to 00.
Send:
set time 14:15
Receive:
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
addr dns
Receive:
addr dns 192.168.1.1
set addr dns address
This command sets the domain name server address, where address
consists of four numbers ranging from 0-255 inclusive, separated by ".".
s
Thermo Fisher Scientific