Thermo Scientific 5028i Instruction Manual page 161

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
3. Loosen both reel nuts on tape spindles. Remove used filter tape and
empty tape spool from tape spindles.
4. Attach end of tape to blank tape spool on left tape spindle with
2-inch of tape for Channel A(Upper), and attach end of tape to
blank tape spool on right tape spindle with 2-inch of tape for
Channel B(Under)
5. Insert new blank pick up spool on right tape spindle for Channel
A(Upper), and insert new blank pick up spool on left tape spindle
for Channel B(Under).
6. Route tape according to
7. Attach end of tape to blank tape spool on right tape spindle with
2-inch of tape for Channel A(Upper), and attach end of tape to
blank tape spool on left tape spindle with 2-inch of tape for Channel
B(Under).
8. Turn blank tape spool until there is two complete wrappings of tape.
9. Tighten right reel nut for Channel A(Upper), and tighten left reel
nut for Channel B(Under). From Filter Tape Control > Manual >
choose Tape and press
10. From the main menu choose Alarms > Instrument Alarms to verify
that there are no filter tape alarms.
11. Confirm the filter tape can move correctly, then close front door.
Figure
5–5.
to MOVE tape.
Model 5028i Instruction Manual
Preventive Maintenance
Filter Tape Replacement
5-7