set alarm flow temp max a 70.0
Receive:
set alarm flow temp max a 70.0 ok
Send:
set alarm flow temp max b 70.0
Receive:
set alarm flow temp max b 70.0 ok
alarm ib status
This command reports the status of the interface board as FAIL or OK.
The following example reports that the interface board status is OK.
Send:
alarm ib status
Receive:
alarm ib status ok
alarm io status
This command reports the status of the I/O expansion board as FAIL or
OK. The following example reports that the I/O expansion board status
is OK.
Send:
alarm io status
Receive:
alarm io status ok
alarm mb status
This command reports the status of the motherboard as FAIL or OK.
The following example reports that the motherboard status is OK.
Send:
alarm mb status
Receive:
alarm mb status ok
alarm sample rh min a/b
alarm sample rh max a/b
These commands report the current sample relative humidity alarm
minimum and maximum value settings. The following example reports
that the sample relative humidity alarm minimum value is 5%.
Send:
alarm sample rh min a
Receive:
alarm sample rh min a 5.0%
Send:
alarm sample rh min b
Receive:
alarm sample rh min b 5.0%
set alarm sample rh min a/b value
set alarm sample rh max a/b value
These commands set the sample relative humidity alarm minimum
and maximum values to value, where value is a floating-point number
representing sample relative humidity alarm limits in percent. The
following example sets the sample relative humidity alarm maximum
value to 40%.
Send:
set alarm sample rh max a 40.0
Receive:
set alarm sample rh max a 40.0 ok
Send:
set alarm sample rh max b 40.0
C-Link Protocol Commands
Model 5028i Instruction Manual
Alarms
B-17