Thermo Scientific 5028i Instruction Manual page 262

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
Diagnostics
B-22
Model 5028i Instruction Manual
sample rh a/b
This command reports the sample RH, in percent. The following
example reports that the sample RH is 20%.
Send:
sample rh a
Receive:
sample rh a 20.0%
Send:
sample rh b
Receive:
sample rh b 20.0%
tape counter a/b
This command reports the number counts which corresponds to the
filter tape movement. The following example reports that the tape count
is 250.
Send:
tape counter a
Receive:
tape counter a 250
Send:
tape counter b
Receive:
tape counter b 250
vac pres a/b
This command reports the current vacuum pressure for the detector, in
mmHg. The following example reports that the vacuum pressure is 100
mmHg.
Send:
vac pres a
Receive:
vac pres a 100.0 mmHg
Send:
vac pres b
Receive:
vac pres b 100.0 mmHg
vac pres offset a/b
This command reports the vacuum pressure offset. The following
example gets the vacuum pressure offset -1.4.
Send:
vac pres offset a
Receive:
vac pres offset a -1.4
Send:
vac pres offset b
Receive:
vac pres offset b -1.4
purge mode a/b
This command get purge mode.The following example get purge mode.
Send:
purge mode a
Receive:
purge mode a 1.000
Send:
purge mode b
Receive:
purge mode b 1.000
set purge mode a/b on
This command sets purge mode on. The following example set purge
mode on.