Thermo Scientific 5028i Instruction Manual page 281

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
set filter period value
This command sets the number of hours to be added to filter next
time for the filter tape change to value, where value is a floating-point
representation of the filter period in hours. The following example sets
the filter period to 10 hours.
Send:
set filter period 10
Receive:
set filter period 10 ok
set filter tape a/b move
This command sets filter tape movement. The following example set
filter tape movement.
Send:
set filter tape a move
Receive:
set filter tape a move ok
Send:
set filter tape b move
Receive:
set filter tape b move ok
filter time
This command reports the next date and time of the filter tape change.
The following example reports that the date and time is January 1,
2008, 6:12 pm.
Send:
filter time
Receive:
filter time 01Jan08 18:12
set filter time dd-mm-yyyy hh:mm:ss
dd = day
mm = month
yyyy = year
hh = hours
mm = minutes
ss = seconds
This command sets the next filter tape change. The following example
sets the filter time to October 1, 2008, 11:32 pm.
Send:
set filter time 10-01-2008 23:32
Receive:
set filter time 10-01-2008 23:32 ok
flow target a/b
This command reports the current flow. The following example reports
that the flow is 16.0 LPM.
Send:
flow target a
Receive:
flow target a 16.00
Send:
flow target b
Receive:
flow target b 16.00
C-Link Protocol Commands
Measurement Configuration
Model 5028i Instruction Manual
B-41

Advertisement

Table of Contents
loading