Thermo Scientific 5028i Instruction Manual page 297

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
set din channel index state
This command assigns digital input channel (1-16) to activate the
action indicated by index (1-12), when the input transitions to the
designated state (high or low). Use the "list var din" command to obtain
the list of supported index values and corresponding actions. The
following example sets the digital input channel 1 to 3 on a low-to-high
transition.
Send:
set din 1 3 high
Receive:
set din 1 3 high ok
dout channel
This command reports the index number, output variable and the active
state assigned to digital output channel. The following example reports
output 4 to be assigned an index number 11 corresponding to action of
"flowl alarms".
Send:
dout 4
Receive:
dout 4 11 FLOW ALARMS open
set dout channel index state
This command assigns digital output channel to be assigned to the
action associated with index, and assigns it an active state of state
(open or closed). Use the "list var dout" command to obtain the list of
supported index values and corresponding state. The following example
sets the digital output channel 4 to state 11.
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
dtoa channel
This command reports the outputs of the 6 or 12 Digital to Analog
converters, according to
the D/A #1 is 97.7% full-scale.
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note If the instrument is in a mode which does not provide a particular
output, and that output is selected, the value will be 0.0.
All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
s
Table
B–13. The following example shows that
Model 5028i Instruction Manual
C-Link Protocol Commands
I/O Configuration
s
B-57