Period - Thermo Scientific 5028i Instruction Manual

Continuous particulate monitor
Table of Contents

Advertisement

START DATE AND TIME:
16 AUG 2016
PRESS
RANGE
Thermo Fisher Scientific
The Next Time screen is used to view and set the initial date and time
Next Time
(24-hour format) for the next filter tape change.
Note Channel B settings please refer to Channel A as following
description.
In the Main Menu, choose Instrument Controls > Filter Tape
l
Control > Channel A or Channel B>Next Time.
00:00
TO EDIT
AVG
DIAGS ALARM
The Period screen allows the user to set the period. The next filter tape

Period

change would be scheduled after the number of hours the period is set
for. The official tape change for TUV and U.S. EPA for PM2.5 and
PM10 is set for 8 hours.
Note Channel B settings please refer to Channel A as following
description.
In the Main Menu, choose Instrument Controls > Filter Tape
l
Control > Channel A orChannel B> Period.
MASS LIMIT:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
s
START DATE AND TIME:
16 AUG 2016
SETTING:
RANGE
AVG
s
PERIOD:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
Instrument Controls Menu
1500
1501
?
MOVE CURSOR
DIAGS ALARM
00:00 ?
DAYS
SET MONTHS
CHANGE VALUE
SAVE VALUE
DIAGS ALARM
24 HRS
000008
HRS?
MOVE CURSOR
SAVE
DIAGS ALARM
Model 5028i Instruction Manual
Operation
3-19