Thermo Scientific 5028i Instruction Manual page 103

Continuous particulate monitor
Table of Contents

Advertisement

Number of Table Points
Thermo Fisher Scientific
The Number of Table Points screen allows the user to select how many
points are used in the analog input conversion table for the selected
channel. The instrument uses linear interpolation between the points
in this table to determine what the reading value is based on the analog
input voltage. Each point in the table consists of an analog input voltage
value (0–10.5 V) and a corresponding reading value. Only two points
are necessary for linear inputs, however a larger number of points may
be used to approximate non-linear inputs. The points range from 2 to
10, with a default of 2.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Input Config > select Channel > Table
Points.
Table Points
The Table Point submenu allows the user to set up an individual table
point.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Input Config > select Channel > Point 1–10.
Volts
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default
table is a two-point table with point 1: 0.00 V = 000.0 U and point 2:
10.00 V = 10.0 U, where U is the previously entered unit of measure.
In the Main Menu, choose Instrument Controls > I/O
l
Configuration > Analog Input Config > select Channel > select
Point > Volts.
NUMBER OF TABLE POINTS:
CURRENTLY:
SET TO:
INC/DEC
SAVE VALUE
RANGE
AVG
DIAGS ALARM
TABLE POINT 01 CONFIG:
>VOLTS
USER VALUE
RANGE
AVG
DIAGS ALARM
TABLE POINT 01 VOLTS:
CURRENTLY:
SET TO:
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
Instrument Controls Menu
2
10 ?
0.00
0.00
0.00
00.00
MOVE CURSOR
SAVE
Model 5028i Instruction Manual
Operation
3-49