Thermo Scientific 5028i Instruction Manual page 115

Continuous particulate monitor
Table of Contents

Advertisement

Min and Max Beta Alarms
RH/Temperature Alarms
Thermo Fisher Scientific
In the Main Menu, choose Alarms > Detector Alarms > Alpha or
l
Alpha B.
The Beat Alarms screen allows the user to view the minimum and
maximum alarm parameters. Acceptable alarm limits range from 5000
to 20000. If the beta count reading goes beyond the maximum limit, an
alarm is activated and the alarm (bell) icon appears in the status bar on
the Run screen and in the Main Menu.
Note Beta B settings please refer to Beta as following description.
In the Main Menu, choose Alarms > Detector Alarms > Beta or
l
Beta B.
The RH/Temperature Alarms submenu allows the user to view and set
the relative humidity or temperature alarm parameters.
In the Main Menu, choose Alarms > RH/Temperature Alarms.
l
The Ambient RH screen displays the current ambient relative humidity
Ambient RH
value and sets the minimum and maximum alarm limits. Acceptable
alarm limits range from 5 to 95%. If the ambient RH reading goes
beyond the minimum or maximum limit, an alarm is activated and the
alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
ALPHA COUNT:
ACTUAL
MIN
MAX
RANGE
AVG
DIAGS ALARM
BETA COUNT:
ACTUAL
MIN
MAX
RANGE
AVG
DIAGS ALARM
RH/TEMPERATURE ALARMS:
>AMBIENT RH
SAMPLE
RH
SAMPLE
RH
B
AMBIENT TEMP
FLOW TEMP
FLOW TEMP B
BOARD TEMP
RANGE
AVG
DIAGS ALARM
Operation
Alarms Menu
0
0
100
0
5000
20000
OK
OK
OK
OK
OK
OK
OK
Model 5028i Instruction Manual
s
3-61