C-Link Protocol Commands
Measurement Configuration
Measurement
Configuration
B-40
Model 5028i Instruction Manual
conc unit
This command reports the current concentration units (mg/m
3
m
). The following example reports that the concentration units are set
3
to mg/m
.
Send:
conc unit
Receive:
conc unit µg/m3
set conc unit unit
3
unit = | µg/m
| mg/m
This command sets the concentration units to mg/m
following example sets the concentration units to mg/m
Send:
set conc unit mg/m3
Receive:
set conc unit mg/m3 ok
custom range
range = | 1 | 2 | 3 |
This command reports the user-defined value of custom range 1, 2, or
3. The following example reports that custom range 1 is defined to 5.50
3
mg/m
.
Send:
custom 1
Receive:
custom 1 5.500E+00 mg/m3
set custom range range value
set custom 1 value
set custom 2 value
set custom 3 value
These commands are used to set the maximum concentration for any
of the three custom ranges 1, 2, or 3 to range value, where value is a
floating-point number representing concentration in the currently
selected units. The following example sets the custom 1 range to 100.5
3
µg/m
.
set custom 1 100.5 m g/m3
Send:
set custom 1 100.5 m g/m3 ok
Receive:
filter period
This command reports the number of hours to be added to filter next
time for the filter tape change. The following example reports that the
filter period is eight hours.
Send:
filter period
Receive:
filter period 8 HRS