Thermo Scientific 5028i Instruction Manual page 35

Continuous particulate monitor
Table of Contents

Advertisement

Thermo Fisher Scientific
Filter Media
Two Sample Filters: Glass fiber filter tape (width 40 mm; length: 42 m)
Inlet Accessories
TSP
U.S. EPA PM10 Inlets
U.S. EPA PM2.5 Inlet Configurations
PM2.5 VSCC™
PM1.0 SCC
Tripod
Data Output and Input
iPort software to view and change system operation from PC
Outputs: Selectable Voltage, RS232/RS485, TCP/IP, 10 Status Relays, and Power Fail
Indication (standard). 0-20 or 4-20 mA Isolated Current Output (optional)
Inputs: 16 Digital Inputs (standard), Eight 0 to 10 VDC Analog Inputs (optional) 8 User-
Defined Analog Outputs (0 -1 or 0 -5 VDC)
Protocols: C-Link, MODBUS, Geystitech (Bayern-Hessen), ESM Protocol, streaming data,
and NTP (Network Time Protocol) protocols. Simultaneous connections from different
locations over Ethernet.
2 User-Defined Contact Closure Alarm Circuits
4 Averaged Analog Inputs (0–5 VDC) with user-defined conversion to engineering units
Data Storage
Maximum of 190,000 Concentration records with date, time, and flags
Up to 32 types of logged data parameters, increased parameters reduces number of
records
Power Requirements
Model 5028i: 100-120 VAC(5028i-A), 220-240 VAC(5028i-D), 50/60 Hz
Pump: 220 VAC 50/60 Hz, 2.0 A
Maximum Power: 700 Watts(110V); 700 Watts(220V)
Physical Dimensions
W: 18.96" (48.2 cm) x D: 22.23" (56.5 cm) x H: 23.43" (59.5 cm)
Weight: 110VAC: 77lbs. (35 kg), 220VAC: 66lbs.(30kg)
Introduction
Specifications
Model 5028i Instruction Manual
1-9