Thermo Scientific 5028i Instruction Manual page 309

Continuous particulate monitor
Table of Contents

Advertisement

(0x03/0x04) Read Holding
Registers/Read Input
Thermo Fisher Scientific
Response
Field Name
Function
Byte Count
Output Status 2–10
Output Status 11–15
The status of outputs 2–10 is shown as the byte value 0xCD, or binary
1100 1101. Output 10 is the MSB of this byte, and output 2 is the
LSB. By convention, bits within a byte are shown with the MSB to the
left and the LSB to the right. Thus, the outputs in the first byte are '10
through 2', from left to right. In the last data byte, the status of outputs
15-11 is shown as the byte value 0x0A, or binary 0000 1010. Output
15 is in the fifth bit position from the left, and output 11 is the LSB of
this byte. The four remaining high order bits are zero filled.
Read Holding/Input Registers reads the measurement data from the
instrument. Issuing either of these function codes will generate the same
response. These functions read the contents of one or more contiguous
Registers
registers.
These registers are 16 bits each and are organized as shown below. All
of the values are reported as 32-bit IEEE standard 754 floating point
format. This uses 2 sequential registers, least significant 16 bits first.
The request specifies the starting register address and the number of
registers. Registers are addressed starting at zero. Therefore, registers
numbered 1–16 are addressed as 0–15. The register data in the response
message are packed as two bytes per register, with the binary contents
right justified within each byte. For each register, the first byte contains
the high order bits and the second contains the low order bits.
Request
Function Code
Starting Address
Quantity of Registers
(Hex)
0x01
0x03
0xCD
0x0A
1 Byte
0x03 or 0x04
2 Bytes
0x0000 to maximum allowed by instrument
2 Bytes
1 to maximum allowed by instrument
MODBUS Protocol
Function Codes
Model 5028i Instruction Manual
C-5