set amb rh offset value
This command sets the ambient RH offset to value, where value is a
floating-point number representing ambient RH offset in percent. The
following example sets the ambient RH offset to 7%.
Send:
set amb rh offset 7
Receive:
set amb rh offset 7 ok
amb temp offset
This command reports the current ambient temperature offset, in
degrees C. The following example reports that the ambient temperature
offset is 5 °C.
Send:
amb temp offset
Receive:
amb temp offset 5.0 degC
set amb temp offset value
This command sets the ambient temperature offset to value, where value
is a floating-point number representing ambient temperature offset in
degrees C. The following example sets the ambient temperature offset to
3 °C.
Send:
set amb temp offset 3
Receive:
set amb temp offset 3 ok
baro mass coef a/b
This command reports the current barometric mass coefficient. The
following example reports that the barometric mass coefficient is
0.000540.
Send:
baro mass coef a
Receive:
baro mass coef a 0.000540
Send:
baro mass coef b
Receive:
baro mass coef b 0.000540
set baro mass coef a/b value
This command sets the barometric mass coefficient to user-defined
values to value, where value is a floating-point representation of the
coefficient. The following example sets the barometric mass coefficient
to 0.000630.
Send:
set baro mass coef a 0.000630
Receive:
set baro mass coef a 0.000630 ok
Send:
set baro mass coef b 0.000630
Receive:
set baro mass coef b 0.000630 ok