Thermo Scientific 5028i Instruction Manual page 290

Continuous particulate monitor
Table of Contents

Advertisement

C-Link Protocol Commands
Communications Configuration
B-50
Model 5028i Instruction Manual
Send:
set addr nm 255.255.252.0
Receive:
set addr nm 255.255.252.0 ok
addr ntp
This command reports the IP address for the network time protocol
server. See "Network Time Protocol Server" in the Communications
Settings" section of the "Operation" chapter for more information.
Send:
addr ntp
Receive:
addr ntp 10.209.43.237
set addr ntp address
This command sets the network time protocol server address, where
address consists of four numbers ranging from 0-255 inclusive, separated
by ".".
Send:
set addr ntp 10.209.43.237
Receive:
set addr ntp 10.209.43.237 ok
allow mode cmd
This command reports the current allow mode setting: 1 = allow "set
mode local" and "set mode remote" commands; 0 = ignore "set mode
local" or "set mode remote"commands, according to
default value is 0; ignore the commands. The following example shows
that the instrument is configured to ignore "set mode local" or "set
mode remote" commands.
Send:
allow mode cmd
Receive:
allow mode cmd 0
set allow mode cmd value
This command is used to configure the instrument to value, where value
is either 1 = accept or 0 = ignore the "set mode local" and "set mode
remote" commands, according to
If the instrument is set to accept the commands (value = 1), the "set
mode local" command will unlock the instrument and the keypad can
be used to make changes via the front panel.
If the instrument is set to ignore the commands (value = 0), the
instrument will respond with "ok" as if the command has been accepted
and acted upon, but will not change the instrument lock status (this is
for compatibility with systems expecting and "ok" response).
Note The instrument will always respond to the command "mode"
with the status of the password lock as "mode local" or "mode remote",
regardless of the above setting.
The following example sets the instrument to accept the "set mode
local" and "set mode remote" commands.
Table
B–7.
s
Table
B–7. The