Auto Mass Coefficient

Manual Mass Coefficient

Alpha Efficiency Coefficient

3-78
Model 5028i Instruction Manual
The Auto Mass Coefficient screen allows the user to conduct a mass foil
calibration.
Note After pressing
mass foil calibration procedure. For details, see
on page
4-8
Bookmark not defined.
In the Main Menu, choose Service > Mass Calibration > Mass
l
Calibration or Mass Calibration B> Mass > Auto.
MASS COEF:
FOIL VALUE:
AND OPEN BENCH
RANGE
The Manual Mass Coefficient screen allows the user to adjust the mass
coefficient used in beta attenuation.
In the Main Menu, choose Service > Mass Calibration > Mass
l
Calibration or Mass Calibration B> Mass > Manual.
MASS COEFFICIENT:
CURRENTLY:
RANGE
The Alpha Efficiency Coefficient screen allows the user to view and set
the alpha efficiency value.
Note This value is set in the factory and should not be adjusted.
In the Main Menu, choose Service > Mass Calibration > Mass
l
Calibration or Mass Calibration B> Alpha Eff Coefficient.
ALPHA EFFICIENCY COEF:
CURRENTLY:
RANGE
a series of screens will walk you through the
s
7000.0
1442
TO SAVE FOIL VALUE
AVG
DIAGS ALARM
SET TO:
MOVE CURSOR
CHANGE VALUE
AVG
DIAGS ALARM
0.120000
SET TO:
0.120001
MOVE CURSOR
CHANGE VALUE
AVG
DIAGS ALARM
"Auto Mass
Coefficient"
µg
3
7000
6992
?
SAVE
s
?