Send:
set allow mode cmd 1
Receive:
set allow mode cmd 1 ok
Table B–7. Allow Mode Command Values
Value
Allow Mode Command
0
Ignore (default)
1
Accept
baud
This command reports the current baud rate for the serial port (RS232/
RS485). The following example reports that the current baud rate is
9600 baud.
Send:
baud
Receive:
baud 9600
set baud rate
rate = | 1200 | 2400 | 4800 | 9600 | 19200 | 38400 | 57600 | 115200 |
This command sets the instrument baud rate to rate. The following
example sets the instrument's baud rate to 9600.
Note After the command is sent, the baud rate of the sending device
must be changed to agree with the instrument.
Send:
set baud 9600
Receive:
set baud 9600 ok
dhcp
This command reports the current state of use of the Dynamic Host
Configuration Protocol (DHCP) as on or off. DHCP is used to assign
an IP address to the instrument automatically. The following example
shows that DHCP is on.
Send:
dhcp
Receive:
dhcp on
set dhcp onoff
onoff = | on | off |
This command enables (on) and disables (off) the DHCP service. When
DHCP is set to on, the instrument gets the IP address, the netmask
address, and the gateway address from a DHCP server. When DHCP is
set to off, the instrument gets these addresses from system memory. The
following example sets the DHCP service to on.