Calibrate Barometer Pressure Span; Restore Default Calibration - Thermo Scientific 5028i Instruction Manual

Continuous particulate monitor
Table of Contents

Advertisement

Calibrate Barometer Pressure

Restore Default Calibration

RESTORE DEFAULT CAL:
RANGE
Vacuum/Flow or Vacuum/ Flow B
Thermo Fisher Scientific
The Calibrate Barometer Pressure Span screen allows the user to view
Span
and pressure sensor calibration span point.
Note The operator should use an independent barometer to measure
the ambient pressure and enter the value on this screen before
calibrating.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
l
Baro Pres Calibration > Span.
The Restore Default Calibration screen allows the user to reset the
default values.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
l
Baro Pres Calibration > Set Defaults.
RESTORE
AVG
DIAGS ALARM
The Vacuum/Flow or Vaccum/Flow B Calibration submenu are used
to calibrate the vacuum/flow zero, vacuum/flow span, or reset default
Calibration
values.
Note Vac/Flow B settings please refer to Vac/Flow as following
description.
In the Main Menu, choose Service > Pres/Vacuum Calibration >
l
Vac/Flow or Vac/Flow B.
s
CAL BARO PRESSURE SPAN:
PRESSURE:
SET TO:
CHANGE VALUE
RANGE
AVG
RESTORE DEFAULT CAL:
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
RANGE
AVG
s
VAC PRES CALIBRATION:
>VAC/FLOW OFFSET
VAC
PRES SPAN
FLOW PRES SPAN
SET DEFAULTS
RANGE
AVG
760 mmHg
747
mmHg?
MOVE CURSOR
SAVE
DIAGS ALARM
RESTORE
TO?x
DIAGS ALARM
1.0000
1.0000
DIAGS ALARM
Model 5028i Instruction Manual
Operation
Service Menu
3-73