Troubleshooting
Troubleshooting Guides
6-4
Model 5028i Instruction Manual
Table 6–2. Troubleshooting - Alarm Messages
Alarm Message
Instrument Alarms – Filter
Tape Counter
Instrument Alarms – Filter
Tape Change
Instrument Alarms –
Bench
Instrument Alarms-Heater
Power
Instrument Alarms –
Motherboard Status
Instrument Alarms –
Interface Status
Instrument Alarms – I/O
Exp Status
Detector Alarms –Board
Status
Detector Alarms – Alpha
Detector Alarms – Beta
Possible Cause
Action
Loose reel nuts
Tighten reel nuts.
Broken tape
Replace tape.
Loose photo interrupt
Verify connections.
cable
Loose reel nuts
Tighten reel nuts.
Broken tape
Replace tape.
Tape motor nut
Verify connection to measurement
turning
interface board.
Replace tape motor.
Cam working not
Verify connector to measurement
functioning
interface board.
Replace motor.
Motor turning, cam
Verify setscrews on cam coupling.
not
Loose heater power
Verify connections.
cable
Measurement
Replace the measurement interface
interface board dose
board.
not work properly
Internal cables not
Check that all internal cables are
connected properly
connected properly. Re-cycle AC
power to instrument. If still alarming,
change board.
Board defective
Replace board.
Board defective
Replace board.
Board defective
Replace board.
Detector
Re-calibrate detector.
Settings
Verify alarm settings are between 0
minimum and 100 maximum.
Disconnected cable
Verify cable connections.
Defective detector
Replace detector-Beta counts
trending towards zero.
Settings
Verify alarm settings are between
5000 minimum and 20000 maximum.
Thermo Fisher Scientific