Maintenance and Calibration

Removing the Mass Transducer

Removing the
Mass Transducer
To remove the mass
5-2
PDM 3600 Instruction Manual
Several maintenance procedure, including changing the filter, cleaning the
grit pot and cleaning the sample lines, require the mass transducer to be
removed from the unit.
1. Locate the TE handle on the left side of the PDM and slide it all the
transducer:
way to the right to unlock the mass transducer
Handle
Figure 5–1. Sliding the handle right to release the mass transducer
2. Pull the mass transducer out of the unit
Figure 5–2. Removing the mass transducer
(Figure
5–1).
(Figure
5–2).