Operation Method (cont'd)
Hints
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
5. Perform the necessary measurement steps for the
selected value using the Trackball and Set key.
When the Set key is pressed to complete the
re-measurement:
S
The resultant value of the selected item and all
related calculations will be recalculated and
displayed on the screen.
S
The new result will be entered on the report page.
S
The old measurement is deleted from the screen
and report page as it is replaced by the new value.
If additional re-measurements need to be performed, repeat the
previous steps.
Each value of an M-Mode continuous sequence can be
remeasured separately.
If a measurement is selected from the soft-menu during the
re-measure process, the re-measurement process is cancelled
and the old value is restored.
After a re-measurement is complete (fixed), the old graphic is
erased and the same cursor symbols are used for the
remeasured value.
If two or more identical measurement items are displayed in the
calculation result area, only the last one supports
re-measurement.
The re-measurement of Heart Rate is always manual, even
though Auto Heart Rate is selected as a preset.
There is no automatic compare function for Systole/Diastole.
It is not possible to do re-measurements on VCR playback
values in Image Memory.
Re-measurement is not possible during a calculation sequence.
Only after the sequence is compiled.
Once cleared or erased, measurements cannot be recalled for
re-measurement.
Advanced Cardiac Calculations
4–39