3D–Surface Mode
Standard Procedure (cont'd)
3–14
3. From the Cine Sub-Menu, set the START FRAME and
END FRAME for the desired images.
At this time, 3D-Surface utilizes all of Cine memory
(When 3DvieW is active, half of the memory is used for
3Dview and the other half is for Cine memory. With
3D-Surface, 3DvieW is not active, so all of Cine
memory is available.)
4. Select the Opacity setting. The available selections
are:
AUTO-1
The standard setting.
AUTO-2
Lower contrast image displayed than
AUTO-1.
AUTO-3
Higher contrast image displayed than
AUTO-1.
5. Press SURFACE ROI. The ROI box cursor is displayed
on the B-Mode image and the Cine Loop starts
automatically.
6. After reviewing the selected CINE images, use the
Trackball and Scan Area key to adjust the box cursor
to the desired size and position for the 3D-Surface
image.
7. To cancel the ROI box without rendering, press
SURFACE ROI. The SURFACE ROI sub-menu
selection is not highlighted, the ROI box cursor
disappears and the Cine Loop stops.
8. Pressing Set, the following message is displayed:
In progress, please wait.
After a while, a temporary 3D-Surface image is
displayed.
9. Depending on the probe movement used to collect the
images, adjust the length/width scale with the Ellipse
rocker switch.
10. Press Set, Clear or SURFACE ROI to fix the
3D-Surface Image.