GE LOGIQ 500 Advanced Reference Manual page 121

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

LV Calculation Formulas (Gibson Method) (cont'd)
Calc Mnemonic
EdV
End Diastole Volume
EsV
End Systole Volume
FS
Fractional Shortening
SV
Stroke Volume
EF
Ejection Fraction
CO
Cardiac Output
SI
Stroke Volume Index
CI
Cardiac Index
MVCF
Mean Vcf
LVM
Left Ventricle
Cardiac Mass
Table 4–11. Gibson Method Calculation Formulas
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
Calc Name
Input Measurements
three distances
three distances
two distances
two distances
two distances
two distances and one 2
beat time interval
body surface area and SV
body surface area and CO
two distances & one time
interval
three distances
Advanced Cardiac Calculations
EdV[ml]=π/6 x (LVIDd)^2 x
(0.98 x LVIDd + 0.59)
EsV[ml]=π/6 x (LVIDs)^2 x
(1.14 x LVIDs + 4.18)
FS=(1–LVIDs/LVIDd)x100
SV[ml]=EdV–EsV
EF=SV/EdVx100
CO[1/min]=SVxHR/1000
SI[ml/m^2]=SV/BSA
CI[ml/min/m^2]=CO/BSA
MVCF[circ/s]=
(LVIDd-LVIDs)/
(LVIDd x ET)
LVM[g]=1.04*[(IVSd+
LVPWd+LVIDd)^3–
LVIDd^3]–13.6
Formula
4–75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents