Standard Procedure (cont'd)
Orientation
Marking
CAUTION
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
To avoid confusion with the 3D-Surface image, ensure that:
S
The orientation mark on the probe and the displayed image
are properly aligned.
Figure 3–7. Proper Probe Orientation
When acquiring the image data, pull the probe from Start to
Finish. DO NOT PUSH THE PROBE.
Pushing the probe may cause the 3D-Surface image to appear
confusing or improper.
3D–Surface Mode
Orientation
Marking
3–13