ULTRASOUND PROGRAM MANAGEMENT GROUP
Date : August 28, 2000
To : Holders of P9030TA LOGIQ 500 Service Manual
Subject : P9030TA LOGIQ 500 SERVICE MANUAL UPGRADE – REV 14
Enclosed please find the following Rev14 upgrade pages.
SUMMARY OF CHANGES (Reason)
Chapter 1: Addition of the caution label and change of address
Chapter 3: Additional information for Ver. 6 system and the new probes
Chapter 4: Additional information for Ver. 6 system software options
Chapter 5: Additional information for Ver. 6 system
Chapter 6: Additional descriptions for new FRUs and others
UPDATE INSTRUCTIONS
To properly upgrade your manual, exchange the upgraded pages in the list below:
CHAPTER
1
3
4
5
6
Yuji Kato
US PROGRAM MANAGEMENT GROUP – ULTRASOUND BUSINESS DIVISION, GEYMS
UPDATE INSTRUCTIONS
REMOVE PAGE
SHEETS
NUMBERS
(Pgs)
Title page REV 13
1
A REV 13
1
i to vi
3
1–1 to 1–2
1
1–9 to 1–20
6
3–1 to 3–2
1
3–11 to 3–16
3
4–1 to 4–2
1
4–15 to 4–16
1
4–21 to 4–26
3
4–35 to 4–40
3
5–11 to 5–12
1
6–1 to 6–252
126
TOTAL
151
REMOVED
INSERT PAGE
SHEETS
NUMBERS
(Pgs)
Title page REV 14
1
A REV 14
1
i to vi
3
1–1 to 1–2
1
1–9 to 1–22
7
3–1 to 3–2
1
3–11 to 3–18
4
4–1 to 4–2
1
4–15 to 4–16
1
4–21 to 4–26
3
4–35 to 4–42
4
5–11 to 5–12
1
6–1 to 6–266
133
TOTAL
161
INSERTED

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ 500

  • Page 1 UPDATE INSTRUCTIONS Date : August 28, 2000 To : Holders of P9030TA LOGIQ 500 Service Manual Subject : P9030TA LOGIQ 500 SERVICE MANUAL UPGRADE – REV 14 Enclosed please find the following Rev14 upgrade pages. SUMMARY OF CHANGES (Reason) •...
  • Page 2 P9030TA Revision 14 LOGIQ 500 Service Manual  Copyright 1994, 1995, 1996, 1997, 1998, 1999, 2000 by General Electric Company...
  • Page 3 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA LIST OF EFFECTIVE PAGES DATE PRIMARY REASON FOR CHANGE March 10, 1994 Initial release July 25, 1994 Software version 1.10 Release, Error correction and Additional descriptions August 15, 1994 Error Correction October 26, 1994 Software version 1.11 Release, Error correction and Additional descriptions...
  • Page 4: Table Of Contents

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA TABLE OF CONTENTS SECTION TITLE PAGE TABLE OF CONTENTS ..............
  • Page 5 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA SECTION TITLE PAGE CHAPTER 3 – SYSTEM CONFIGURATION 3–1 INTRODUCTION ..............
  • Page 6 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA SECTION TITLE PAGE CHAPTER 5 – DIAGRAM 5–1 INTRODUCTION ..............
  • Page 7 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA SECTION TITLE PAGE CHAPTER 6 – RENEWAL PARTS (continued) 6–2–28 Keyboard Bottom Cover (FRU No. 309) ........
  • Page 8 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA SECTION TITLE PAGE CHAPTER 6 – RENEWAL PARTS (continued) 6–2–71 HDDB Assy (FRU No. 522) ..........
  • Page 9 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA SECTION TITLE PAGE CHAPTER 6 – RENEWAL PARTS (continued) 6–6 REPLACING 2.5–INCH HDD ASSY WITH 3.5–INCH HDD ASSY ......
  • Page 10 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA CHAPTER 1 – INTRODUCTION TABLE OF CONTENTS SECTION TITLE PAGE 1–1 SERVICE MANUAL CONTENTS ............
  • Page 11 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA This page is left blank intentionally. 1–2 INTRODUCTION...
  • Page 12 Chapter 5, Diagrams: Contains block diagrams and functional explanations of the LOGIQ 500 electronics; Chapter 6, Renewal Parts: Contains a complete list of replacement parts for the LOGIQ 500 and disassembly procedures for all changeable FRU;...
  • Page 13 WARNING! CAREFULLY READ ALL THE WARNINGS LISTED BELOW! The operator manual should be fully read and understood before operating the LOGIQ 500 and kept nearby for quick reference. Although the ultrasound energy transmitted from the LOGIQ 500 transducer is within AIUM/NEMA standards, unnecessary exposure should be avoided.
  • Page 14 P9030TA 1–2–1 Warnings (continued) [Consoles with Software prior to V5.00 y] [Consoles with V5.00 y or Later] OUTSIDE MARKINGS OF LOGIQ 500 (For All Units) ILLUSTRATION 1–1 Note This caution label has been removed at the installation of the system.
  • Page 15 1–2–1 Warnings (continued) [Consoles with Software prior to V5.00 y] [Consoles with V5.00 y or Later] OUTSIDE MARKINGS OF LOGIQ 500 (For All Units) ILLUSTRATION 1–2 Note For further details regarding the cautions above, refer to 2–2–8 MOVING INTO POSITION in Chap- ter 2.
  • Page 16 Do not place any objects on the monitor. [Consoles with Software prior to V5.00 y] [Consoles with V5.00 y or Later] OUTSIDE MARKINGS OF LOGIQ 500 (For All Units) ILLUSTRATION 1–3 1–7...
  • Page 17 LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 1–2–1 Warnings (continued) OUTSIDE MARKINGS OF LOGIQ 500 (For USA/Asia) ILLUSTRATION 1–4 Note The ETL label described in ILLUSTRATION 1–4 is attached on the system software version 2.00 or later console. The system software 1.20 or former console has the different style of ETL label.
  • Page 18 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 1–2–1 Warnings (continued) OUTSIDE MARKINGS OF LOGIQ 500 (For Europe) [a] ILLUSTRATION 1–5 Note For the symbols shown in the illustration above, refer to latter pages in this chapter. 1–9...
  • Page 19 LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–2–1 Warnings (continued) OUTSIDE MARKINGS OF LOGIQ 500 (For Europe) [b] ILLUSTRATION 1–6 Note The ILLUSTRATION 1–6 shows the labels for console with the software version 2.00 or later. The ILLUSTRATION 1–5 shows the labels for console with the software version 1.20 or former.
  • Page 20 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–2–1 Warnings (continued) CE Marking Label (For Europe) ILLUSTRATION 1–7 Note The ILLUSTRATION 1–7 shows the labels for console with the software version 6.00 or later.The LO- GIQ 500 system conforms to the CISPR11, Group 1, Class A of the international standard for Electromagnetic disturbance characteristics.
  • Page 21: Chapter 1 - Introduction

    1–2–1 Warnings (continued) Labels including English, Russian, Swedish, Danish, Greek, and Turkish [supplied with European Console] OUTSIDE MARKINGS OF LOGIQ 500 (For Europe) [c] ILLUSTRATION 1–8 Note The labels shown in ILLUSTRATION 1–8 are supplied with the consoles for Europe. They shall be attached on the console over the existing labels as necessary.
  • Page 22 LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–2–1 Warnings (continued) OUTSIDE MARKINGS OF LOGIQ 500 (For Japan) ILLUSTRATION 1–9 Note For the symbols shown in the illustration above, refer to latter pages in this chapter. The CAUTION label for the radio influence is attached on the console from April, 1996.
  • Page 23 CAUTION Do not use a Defibrillator simultaneously with the ECG, as its excessive voltage will damage the signal input block of the ECG unit. OUTSIDE MARKINGS OF LOGIQ 500 (For Units with ECG) ILLUSTRATION 1–10 Note This label is attached only on the LOGIQ 500 console with the optional ECG unit.
  • Page 24 1–2–1 Warnings (continued) NOTE: Same labels are attached on both left and right side of the transformer box inside the side covers MARKINGS OF LOGIQ 500 (INSIDE SIDE COVERS) ILLUSTRATION 1–12 Note For further details, refer to 4–4 POWER SUPPLY ADJUSTMENT in Chapter 4.
  • Page 25 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–2–1 Warnings (Continued) The following table describes the purpose and location of safety labels and other important information provided on the equipment. Label/Symbol Purpose/Meaning Location • Manufacturer’s name and address...
  • Page 26 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–2–1 Warnings (Continued) Label/Symbol Purpose/Meaning Location • ”CAUTION – Dangerous voltage” Various (the lightning flash with arrowhead in equilateral triangle) is used to indi- cate electric shock hazards. • ”Mains OFF” Indicates the power...
  • Page 27: Specifications

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–2–2 Specifications Type of protection against electric shock: Class I EQUIPMENT (*1) Degree of protection against electric shock: Type BF EQUIPMENT (*2) (Except ECG) Type CF EQUIPMENT (*3) (ECG Only)
  • Page 28: Emc (Electromagnetic Compatibility)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–3 EMC (Electromagnetic Compatibility) 1–3–1 EMC Performance All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, ei- ther through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates capability of the equip- ment, which curbs electromagnetic influence from other equipment and at the same time does not affect other equip- ment with similar electromagnetic radiation from itself.
  • Page 29: General Notice

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 1–3–3 General Notice 1) Designation of Peripheral Equipment Connectable to This Product The equipment which conforms to EM60601–1–2 (IEC601–1–2), can be hooked up to the product without compromising its EMC performance.
  • Page 30: Address

    (1) 800–321–7937 FAX: (1) 414–647–4125 LATIN & SOUTH AMERICA TEL: (1) 414–524–5300 GE Ultrasound Europe GE Ultraschall Deutschland GmbH & Co. KG Beethovenstr. 239 42655 Solingen, GERMANY TEL: OLC–Europe Toll Free Numbers General Imaging Hotline (49) (212) 2802 207 Cardiac Hotline...
  • Page 31 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 1–22 INTRODUCTION...
  • Page 32 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA CHAPTER 2 – INSTALLATION TABLE OF CONTENTS SECTION TITLE PAGE 2–1 PREINSTALLATION ..............
  • Page 33 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA This page is left blank intentionally. 2–2 INSTALLATION...
  • Page 34 2–1–1 Introduction This section describes various general electrical, operational, and environmental considerations that must be consid- ered before installing the LOGIQ 500 Ultrasound unit. 2–1–2 Power Line Requirements The following power line parameters should be monitored for one week before installation. We recommend that you use an analyzer Dranetz Model 606–3 or Dranetz Model 626:...
  • Page 35 Operating Conditions The LOGIQ 500 is designed to operate within a temperature range of 10 degrees C to 40 degrees C (50 degrees F to 104 degrees F), and a relative humidity range of 30 % to 85 % (Non–condensing).
  • Page 36 TEMPERATURE 20–26 DEG. C (68–79 DEG F) for PATIENT COMFORT HUMIDITY 50% to 70% for PATIENT COMFORT HEAT DISSIPATION 2000 BTU/Hr for LOGIQ 500 ; FLOOR LOADING Approximately 680 – 800 kg/m without Accessories FLOOR CONDITION Gradient : WITHIN 5 degrees...
  • Page 37: Recommended Ultrasound Room Layout

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 2–1–4 Recommended Ultrasound Room Layout (Continued) Optional Desirable Ultrasound Room Facilities These facilities are: Lab sink with hot and cold water; Emergency oxygen supply; Dimmer control for overhead lights; Film viewer;...
  • Page 38: Recommended Ultrasound Room Layout

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 2–1–4 Recommended Ultrasound Room Layout (continued) RECOMMENDED ULTRASOUND FLOOR PLAN ILLUSTRATION 2–1 2–7 INSTALLATION...
  • Page 39: Installation

    This section contains many of the procedures required to install the LOGIQ 500 console. 2–2–2 Average Installation Time The LOGIQ 500 has been designed to be installed and checked out by an experienced service technician in approxi- mately four hours. LOGIQ 500 consoles with optional equipment may take slightly longer.
  • Page 40 Remove the SLOPE. Remove the UPPER and LOWER CUSHIONs. Unlock the brakes by stepping down on the brake pads in front, then carefully roll the LOGIQ 500 rear side first off the SKID. 10. Remove the Caution Label attached on the CRT Filter and clean the CRT Filter.
  • Page 41: Unpacking Logiq 500

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 2–2–5 Unpacking LOGIQ 500 (Continued) UNPACKING LOGIQ 500 ILLUSTRATION 2–2 2–10 INSTALLATION...
  • Page 42: Mtz Probe Holder Installation

    ILLUSTRATION 2–3 2–2–6 MTZ Probe Holder Installation One MTZ probe holder is supplied with the LOGIQ 500 console. Assemble the MTZ probe holder at the bottom of standard probe holder by screwing two screws ( 1 and 2 : supplied with the starter kit) as shown in ILLUSTRATION 2–4.
  • Page 43: Mtz Probe Holder Installation

    2–2–6 MTZ Probe Holder Installation (Continued) The standard probe holder is changed to new one from the LOGIQ 500 console with system software version 2.00. Only one MTZ probe holder can be installed on the console with the new standard probe holder. Refer to ILLUSTRATION 2–5.
  • Page 44: Moving Into Position

    Do not lift the console by holding any covers. In general, a single adult can move the LOGIQ 500 along an even surface with no steep grades. At least two people should move the machine when large humps, grooves, or grades will be encountered. (It is better to pull from the rear rather than push from the front of the unit).
  • Page 45: Product Locator Installation Card

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 7 P9030TA 2–2–10 Product Locator Installation Card Fill out proper customer Information the Product Locator Installation Card. Refer to ILLUSTRATION 2–6. Mail this Installation Card “Product Locator” to the address corresponding to your pole.
  • Page 46 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA CHAPTER 3 – SYSTEM CONFIGURATION TABLE OF CONTENTS SECTION TITLE PAGE 3–1 INTRODUCTION ..............
  • Page 47 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA This page is left blank intentionally. 3–2 SYSTEM CONFIGURATION...
  • Page 48: Introduction

    This chapter describes system configuration and specifications. 3–2 DIMENSIONS Regarding LOGIQ 500 dimensions, Refer to ILLUSTRATION 3–1 for planning the position of your LOGIQ 500. Note The weight of the console with software version 5.00y or later is approximately 192 kg (423 lbs).
  • Page 49: Dimensions

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 3–2 DIMENSIONS (continued) ILLUSTRATION 3–2 shows the dimensions for the LOGIQ 500 Tall type console with the software version 5.00 y or later. Note The tall type console is 75 mm taller than the normal height console.
  • Page 50: Electrical Specifications

    3–4 STORAGE AND OPERATION REQUIREMENTS The LOGIQ 500 is shipped in a single container excluding PROBES. Shipping weight is approximately 472 lbs (214kg). The size of the container is B120 cm x W66 cm x H139 cm. (47 in. x 26 in. x 55 in). Table 3–1 provides a summary of temperature, atmospheric pressure, and humidity tolerances for shipping, installation, and operation: TABLE 3–1...
  • Page 51: Optional Peripherals

    3–5 OPTIONAL PERIPHERALS 3–5–1 Peripherals/Accessories Connector Panel LOGIQ 500 peripherals and accessories can be properly connected using the rear connector panel located behind the rear door. Located on the panel are video input and output connectors, audio input and output, camera expose connectors, foot- switch connector power connector and control connections for VCR, printer, MIC and service tools.
  • Page 52 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 3–5–1 Peripherals/Accessories Connector Panel (continued) Note The SCSI port is not available for the console with the 3.5–inch HDD. The SCSI Terminator is not supplied with the console with the 3.5–inch HDD.
  • Page 53 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 3–5–1 Peripherals/Accessories Connector Panel (continued) Note Output level of RS232C signals: High +3V to +15V –15V to 3–8 SYSTEM CONFIGURATION...
  • Page 54 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 3–5–1 Peripherals/Accessories Connector Panel (continued) Note The SCSI Bus connector and the SCSI Terminator plug are not supplied for the console with the Rear CONN Panel2 Assy, 2211163, 2211164,and 2211165.
  • Page 55 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 3–5–1 Peripherals/Accessories Connector Panel (continued) Note Output level of control signals indicated in the above tables are TTL level. 3–10 SYSTEM CONFIGURATION...
  • Page 56: List Of Optional Peripherals

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 3–5–2 List of Optional Peripherals The tables below shows the suggested optional peripherals for LOGIQ 500. 1. RECORDING DEVICES TABLE 3–2 LIST OF RECORDING DEVICES DEVICE MANUFACTURER MODEL VIDEO SIGNAL...
  • Page 57 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 3–5–2 List of Optional Peripherals (continued) 2. TRANSDUCER (PROBE) TABLE 3–3 LIST OF TRANSDUCERS WITH PROBE MATERIAL OF AREA OF CATALOG REQUIRED TYPE FAMILY PART No. FERRITE NAME HEADSHELL USING...
  • Page 58 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 3–5–2 List of Optional Peripherals (continued) 2. TRANSDUCER (PROBE) TABLE 3–3 LIST OF TRANSDUCERS (continued) WITH PROBE MATERIAL OF AREA OF CATALOG REQUIRED TYPE FAMILY PART No. FERRITE NAME HEADSHELL...
  • Page 59 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 3–5–2 List of Optional Peripherals (continued) 2. TRANSDUCER (PROBE) (continued) TABLE 3–3 LIST OF TRANSDUCERS (continued) WITH PROBE MATERIAL OF AREA OF CATALOG REQUIRED TYPE FAMILY PART No. FERRITE NAME...
  • Page 60 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 3–5–2 List of Optional Peripherals (continued) 2. TRANSDUCER (PROBE) (continued) Note 546L, S317, S611, C386, I739 and T739 probes are available on consoles from software version 2.20. Note LA39 and S222 probes are available on consoles from software version 3.00.
  • Page 61 3–5–2 List of Optional Peripherals (continued) 3. CONNECTING CABLES CAUTION Equipment damage possibility. Be sure to use the following recommended connecting cables to connect recording devices and a network with LOGIQ 500 console. TABLE 3–4 LIST OF CONNECTING CABLES NAME PART NO.
  • Page 62: Power Consumption Of Optional Peripherals

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 3–5–3 Power Consumption of Optional Peripherals The table below shows the power consumption of each optional peripheral for LOGIQ 500. TABLE 3–5 POWER CONSUMPTION OF OPTIONAL RECORDING DEVICES DEVICE MODEL...
  • Page 63 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 3–6 VIDEO SPECIFICATIONS TABLE 3–6 VIDEO SPECIFICATIONS 12–inch monitor 15–inch monitor NTSC NTSC Total Line per Frame [Line] Vertical Frame Frequency [Hz] Horizontal Scanning Frequency [kHz] 15.734 15.625 31.469 31.250...
  • Page 64: Power Supply

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA CHAPTER 4 – FUNCTIONAL CHECKS TABLE OF CONTENTS SECTION TITLE PAGE 4–1 INTRODUCTION ..............
  • Page 65 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA This page is left blank intentionally. 4–2 FUNCTIONAL CHECKS...
  • Page 66: Chapter 4 - Functional Checks

    REV 0 P9030TA 4–1 INTRODUCTION This chapter provides procedures for quickly checking major functions of the LOGIQ 500 console, diagnostics by using the built–in service software, and power supply adjustments. 4–1–1 Required Equipment To perform these tests, you’ll need any of a sector, linear, or a convex transducer.
  • Page 67: Basic Controls

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–2–1 Basic Controls Step Check Expected result Connect the sector or convex transducer to ”Probe 1” connector beneath keyboard. Turn ON Power toggle switch. Type USER ID and then press PATIENT ENTRY MENU should appear.
  • Page 68: Basic Controls

    Edge Enhancement up/down. M–MODE DISPLAY SCREEN ILLUSTRATION 4–2 Note You can select several types of display format by using the Preset Menu. For the Preset Menu, refer to Customizing Your System in the LOGIQ 500 User Manual. 4–5 FUNCTIONAL CHECKS...
  • Page 69 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–2–1 Basic Controls (Continued) Step Check Expected result Press CFM key. The CFM mode screen should appear as shown in ILLUSTRATION 4–3. Rotate Gain knob Color image noise should appear.
  • Page 70 You can select the two types of display format: Side by Side (with the 2–D display to the left of the PD–mode display) and Top/Bottom (with the 2–D display on top of the PD–mode display) by using the Preset Menu. For the Preset Menu, refer to Customizing Your System in the LOGIQ 500 User Manual. 4–7...
  • Page 71 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–2–1 Basic Controls (Continued) Step Check Expected result (If the sector probe is The CWD mode screen should appear as shown in connected) ILLUSTRATION 4–5. Press CWD key. CWD MODE DISPLAY SCREEN ILLUSTRATION 4–5...
  • Page 72: Diagnostics

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–3 DIAGNOSTICS 4–3–1 Service Software Menu Select the Set Up function by pressing the Top Menu Select button located above the Set Up section as shown in ILLUSTRATION 4–6. MENU PANEL ILLUSTRATION 4–6...
  • Page 73 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–3–2 Diagnosis Test Menu Pressing the position of the Sub–menu key located under the ECG section selects User Diagnostics function and the Diagnosis Menu is displayed on the CRT monitor as shown in ILLUSTRATION 4–7.
  • Page 74 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 8 P9030TA 4–3–2 Diagnosis Test Menu (Continued) (A) SYSTEM TEST 1 Be sure to connect any probes with the unit to check the following system test. Type 1 and then press Enter key when the Diagnosis Menu is displayed. The system test will start automatically.
  • Page 75 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–3–2 Diagnosis Test Menu (Continued) (B) TEST PATTERN BLACK & WHITE The black & white pattern permits the serviceman to accurately adjust the contrast of an optional equipment (such as a camera) to see the details in the lightest and darkest areas of the image.
  • Page 76 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–3–2 Diagnosis Test Menu (Continued) (C) TEST PATTERN COLOR The color scale permits the serviceman to adjust the color contrast of an optional equipment (such as a color video printer).
  • Page 77 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA 4–3–2 Diagnosis Test Menu (Continued) (D) TEST PATTERN GRAPHICS The vertical linearity and graphics perimeter pattern should be centered on the optional equipment display. This ensures all whitescale graphics overlays will be visible.
  • Page 78 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–3–3 Utility Menu Pressing the position of the Sub–Menu Select key located under the Set Up section selects User Utility function and the Utility Menu is displayed on the CRT monitor as shown in ILLUSTRATION 4–12.
  • Page 79: (A) Time Adjustment

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 5 P9030TA 4–3–3 Utility Menu (Continued) (A) TIME ADJUSTMENT This changes the current time in the system. Type 1 and then press Enter key when the Utility Menu is displayed. The system date and time previously set is displayed as shown in ILLUSTRATION 4–13.
  • Page 80: (B) Error Log Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 4–3–3 Utility Menu (Continued) (B) ERROR LOG DISPLAY This loads and displays an error log file, or saves it into a Floppy disk for the console with software version 1.20 or former or into a MO disk for the console with software version 2.00 or later.
  • Page 81: (C) Trouble Image Save/Load/Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 7 P9030TA 4–3–3 Utility Menu (Continued) (C) TROUBLE IMAGE SAVE/LOAD/DISPLAY This saves the trouble image recorded in the Hard Disk into a Floppy or MO (Magneto Optical) Disk. This also loads the image data saved in the Floppy or MO Disk into the Hard Disk and displays the selected trouble image on the CRT monitor by using the image recall function.
  • Page 82 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 4–3–3 Utility Menu (Continued) (C) TROUBLE IMAGE SAVE/LOAD/DISPLAY (Continued) Procedure for Saving Trouble Images 1. Type 1 and then press Return key. The message “Set a media to drive and then press any key.” is displayed.
  • Page 83 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 4–3–3 Utility Menu (Continued) (C) TROUBLE IMAGE SAVE/LOAD/DISPLAY (Continued) Procedure for Displaying Trouble Image 1. Type 3 and then press Return key. The Trouble Image File Menu will be displayed. Refer to ILLUSTRATION 4–15.
  • Page 84: (D) User Data Backup

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 4–3–3 Utility Menu (Continued) (D) USER DATA BACK UP This saves the user preset data into the external recording media. This also loads the user preset data stored in external recording media to the hard disk.
  • Page 85 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–3–3 Utility Menu (Continued) (D) USER DATA BACK UP (Continued) Type 4 and then press Return key when the Utility Menu is displayed. The User Data Back Up Menu is displayed as shown in ILLUSTRATION 4–16.
  • Page 86 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 4–3–3 Utility Menu (Continued) (D) USER DATA BACK UP (Continued) 3. Input the menu number you desire to save into the disk. Note The system software version 2.00 or later console can save all user data using one MO disk. Type 9 and then press Return key to save all user data into a MO disk.
  • Page 87: (E) User Option Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–3–3 Utility Menu (continued) (E) USER OPTION DISPLAY This displays the option names and the current status for each option. Type 5 and then press Return key when the Utility Menu is displayed. The User Option Display is displayed as shown in ILLUSTRATION 4–18.
  • Page 88 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) (E) USER OPTION DISPLAY (continued) Note The following five (5) options are available only for the console with software version 3.00 or later. Item 25: Multigestation Calculation...
  • Page 89: (F) Board Configuration Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–3–3 Utility Menu (continued) (E) USER OPTION DISPLAY (continued) Note This following three (3) options are available only for the console with the system software version 4.20 or later. Item 38:...
  • Page 90 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 4–3–3 Utility Menu (continued) (F) BOARD CONFIGURATION DISPLAY (continued) BOARD CONFIGURATION DISPLAY (b) ILLUSTRATION 4–19b Note The board configuration and version of each board indicated on board configuration display are differ- ent depending on the style of console.
  • Page 91: (G) Software Configuration Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) (G) SOFTWARE CONFIGURATION DISPLAY This displays the configurations of all system software installed in the console. Type 7 and then press Return key when the Utility Menu is displayed. The Software Configuration Display is displayed as shown in ILLUSTRATION 4–20.
  • Page 92: (H) Modem Set Up

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) (H) MODEM SET UP This displays and changes the current settings for the modem. Type 8 and then press Return key when the Utility Menu is displayed. The Modem Set Up is displayed as shown in ILLUSTRATION 4–21.
  • Page 93: (I) Media Initialize

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 4–3–3 Utility Menu (continued) (I) MEDIA INITIALIZE This initialize the floppy disk. Type 9 and then press Return key when the Utility Menu is displayed. The Media Initialize screen is displayed as shown in ILLUSTRATION 4–22.
  • Page 94: (J) System Id Entry/Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) SYSTEM ID ENTRY/DISPLAY This displays the current system ID number and set the new one. Note This function can be used on the consoles being installed the system software version 2.00 or later.
  • Page 95 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) SYSTEM ID ENTRY/DISPLAY (continued) Procedure for Setting New System ID 1. Type 1 and then press Return key. The system ID number entry menu is displayed as shown in ILLUSTRATION 4–23.
  • Page 96 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) MONITOR ADJUSTMENT This helps to adjust the settings of the Contrast and the Brightness on the CRT Monitor suitably. Note This function is available only for the console with the system software version 4.20 or later.
  • Page 97 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA 4–3–3 Utility Menu (continued) MONITOR ADJUSTMENT (continued) Procedure for Adjusting Monitor 1. Input the item number according to the brightness of Exam room. The Gray Scale Bar for adjustment will be displayed on the CRT monitor screen as shown in ILLUSTRATION 4–25.
  • Page 98 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–3–3 Utility Menu (continued) MONITOR ADJUSTMENT (continued) Procedure for Adjusting Monitor for 15–inch Monitor 1. Press the monitor’s toggle button (2) for brightness and contrast. Confirm that the brightness (or contrast) is displayed on the monitor.
  • Page 99: (K) Network Error Log Display

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–3–3 Utility Menu (continued) NETWORK ERROR LOG DISPLAY This displays and copies an error log file into a MO disk. Note This function is available only when the Storage/Print DICOM option is installed in the console with the system software version 4.10 or later.
  • Page 100 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–4 POWER SUPPLY ADJUSTMENTS CAUTION HAZARDOUS VOLTAGE. 380 Vdc and 253–260 Vac can cause a severe injury or death, or the power supply to be damaged. Turn OFF the power and check the residual voltage of capaci- tors before accessing the power box.
  • Page 101: Power Supply Adjustments

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–4 POWER SUPPLY ADJUSTMENTS (continued) This section provides power supply adjustment procedures for the LOGIQ 500. Adjustments should be only made when necessary. Power supply adjustments should be made in accordance with the schedule for periodic mainte- nance in Chapter 7 of this manual.
  • Page 102 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–4–1 Power Supply Access (continued) Remove five screw caps and unscrew five screws to remove the Side Cover Right as shown in the ILLUSTRATION 4–30. RIGHT COVER REMOVAL ILLUSTRATION 4–30 Unscrew five screws to remove the PS Cover Right.
  • Page 103: Power Supply Adjustment Procedure

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–4–2 Power Supply Adjustment Procedure Power LOGIQ 500 ON. Wait for about 30 seconds to warm up the console. For each of the power supplies which are located at the bottom of LOGIQ 500, connect a DVM to the appropriate place shown in TABLE 4–2.
  • Page 104: Power Supply Adjustment Procedure

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 4–4–2 Power Supply Adjustment Procedure (continued) POWER SUPPLIES ILLUSTRATION 4–32 4–41 FUNCTIONAL CHECKS...
  • Page 105 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 4–42 FUNCTIONAL CHECKS...
  • Page 106: Introduction

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 5 P9030TA CHAPTER 5 – DIAGRAMS TABLE OF CONTENTS SECTION TITLE PAGE 5–1 INTRODUCTION ..............
  • Page 107 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 0 P9030TA This page is left blank intentionally. 5–2 DIAGRAMS...
  • Page 108: Introduction

    SYSTEM The LOGIQ 500 has a 64 channel beamformer that digitizes the RF signal. It uses analog and digital delay to focus the beam. The analog delay is short and fixed. The digital delay is used for long delay and dynamic focus. It offers parallel receive beam formation which can increase frame rate by a factor of two in time critical applications like Color Flow.
  • Page 109: Logiq 500 System

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 7 P9030TA 5–3 BLOCK DIAGRAM LOGIQ 500 SYSTEM BLOCK DIAGRAM ILLUSTRATION 5–1 (a) 5–4 DIAGRAMS...
  • Page 110 NOTE: The DICOM option is available for the console with software ver- sion 4.00 or later. NOTE: VIPP Assy is installed on the con- sole with software version 5.00y or later. VIPP VIPB (or VIDO) LOGIQ 500 SYSTEM BLOCK DIAGRAM ILLUSTRATION 5–1 (b) 5–5 DIAGRAMS...
  • Page 111: Wiring Diagram

    5–4 WIRING DIAGRAM NOTE: The CB46 and CB45N are not installed in the console with the new 3.5–inch HDD (Quantum Fireball 1280S) because the HDD LED Assy is not supplied. LOGIQ 500 SYSTEM WIRING DIAGRAM ILLUSTRATION 5–2 (a) 5–6 DIAGRAMS...
  • Page 112 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 5–4 WIRING DIAGRAM (continued) NOTE: 15–inch Monitor is installed in the console with the software version 5.00y or later. LOGIQ 500 SYSTEM WIRING DIAGRAM ILLUSTRATION 5–2 (b) 5–7 DIAGRAMS...
  • Page 113 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 9 P9030TA 5–5 POWER SUPPLY BLOCK DIAGRAM LOGIQ 500 SYSTEM POWER SUPPLY BLOCK DIAGRAM ILLUSTRATION 5–3 (a) 5–8 DIAGRAMS...
  • Page 114: Power Supply Block Diagram

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 7 P9030TA 5–5 POWER SUPPLY BLOCK DIAGRAM (Continued) LOGIQ 500 SYSTEM POWER SUPPLY BLOCK DIAGRAM ILLUSTRATION 5–3 (b) 5–9 DIAGRAMS...
  • Page 115: Logiq 500 System

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 5–6 CIRCUIT BOARD DESCRIPTION The following table lists circuit boards and their respective card cage slot assignments on the mother board of the LOGIQ 500 system. TABLE 5–1 CIRCUIT BOARD DESCRIPTION...
  • Page 116: Circuit Board Description

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 5–6 CIRCUIT BOARD DESCRIPTION (continued) TABLE 5–2 CIRCUIT BOARD VARIATIONS BOARD NAME VARIATIONS NOTE VERSION BF16 ASSY BF16 ASSY BF16 ASSY can be replaced with BF16–2 ASSY HBBF: 4 HBFC: 5 BF16–2 ASSY...
  • Page 117 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA 5–6 CIRCUIT BOARD DESCRIPTION (continued) TABLE 5–3 CIRCUIT BOARD VARIATIONS (continued) BOARD NAME VERSION VARIATIONS NOTE VIDO ASSY VIDO ASSY VIDO ASSY is replaced with VIDO2 ASSY by FMI72076 VIDO2 ASSY 2117032–2...
  • Page 118: Disassembly/Re-Assembly

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA CHAPTER 6 – RENEWAL PARTS TABLE OF CONTENTS SECTION TITLE PAGE 6–1 RENEWAL PARTS ..............
  • Page 119 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA TABLE OF CONTENTS (continued) SECTION TITLE PAGE 6–2–35 Air Filter (FRU No. 318A), PW Filter (FRU No. 318B) ......
  • Page 120 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA TABLE OF CONTENTS (continued) SECTION TITLE PAGE 6–2–76 Time Keeper RAM (FRU No. 611) ..........
  • Page 121 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 13 P9030TA LIST OF ABBREVIATIONS Stock item Non stock item 6–4 RENEWAL PARTS...
  • Page 122 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 6–1 RENEWAL PARTS LOGIQ 500 MATERIAL LIST (1) PART NAME QUANTITY PART NO. DESCRIPTION OPERATOR CONSOLE ASSY P9522BA 120V, NTSC VER1.11B OPERATOR CONSOLE ASSY P9522CA 220V, PAL VER1.11B OPERATOR CONSOLE ASSY...
  • Page 123 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA LOGIQ 500 MATERIAL LIST (2) PART NAME QUANTITY PART NO. DESCRIPTION OPERATOR CONSOLE ASSY 2126615 120V, NTSC VER2.0 OPERATOR CONSOLE ASSY 2126624 220V, PAL VER2.0 OPERATOR CONSOLE ASSY 2126607 100V, NTSC VER2.0...
  • Page 124 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA LOGIQ 500 MATERIAL LIST (3) PART NAME QUANTITY PART NO. DESCRIPTION OPERATOR CONSOLE ASSY 2149595 120V, NTSC VER2.20 OPERATOR CONSOLE ASSY 2149597 1 1 1 220V, PAL VER2.20 OPERATOR CONSOLE ASSY...
  • Page 125 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA LOGIQ 500 MATERIAL LIST (4) PART NAME QUANTITY PART NO. DESCRIPTION OPERATOR CONSOLE ASSY 2164533 120V, NTSC VER3.00 OPERATOR CONSOLE ASSY 2164534 220V, PAL VER3.00 OPERATOR CONSOLE ASSY 2164531 100V, NTSC VER3.00...
  • Page 126 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA LOGIQ 500 MATERIAL LIST (5) PART NAME QUANTITY PART NO. DESCRIPTION OPERATOR CONSOLE ASSY 2180824 120V, NTSC VER3.10 OPERATOR CONSOLE ASSY 2180825 220V, PAL VER3.10 OPERATOR CONSOLE ASSY 2180823 100V, NTSC VER3.10...
  • Page 127 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA LOGIQ 500 MATERIAL LIST (6) PART NAME QUANTITY PART NO. DESCRIPTION OPERATOR CONSOLE ASSY 2196726 120V, NTSC VER4.00 OPERATOR CONSOLE ASSY 2196727 220V, PAL VER4.00 OPERATOR CONSOLE ASSY 2186725 100V, NTSC VER4.00...
  • Page 128 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA LOGIQ 500 MATERIAL LIST (7) PART NAME PART NO. QUANTITY DESCRIPTION OPERATOR CONSOLE ASSY 2212962 120V, NTSC VER4.10 OPERATOR CONSOLE ASSY 2212963 220V, PAL VER4.10 OPERATOR CONSOLE ASSY 2212961 100V, NTSC VER4.10...
  • Page 129 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA LOGIQ 500 MATERIAL LIST (7) (continued) PART NAME PART NO. QUANTITY DESCRIPTION COLOR ENHANCEMENT S/W 2165092 1 1 1 1 1 MULTIGESTATION S/W 2165093 1 1 1 1 1 ADVANCED VASCULAR...
  • Page 130 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA LOGIQ 500 MATERIAL LIST (8) PART NAME PART NO. QUANTITY DESCRIPTION OPERATOR CONSOLE ASSY 2240875 120V, NTSC VER4.20 OPERATOR CONSOLE ASSY 2242843 220V, PAL VER4.20 OPERATOR CONSOLE ASSY 2240876 220V, PAL VER4.20...
  • Page 131 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA LOGIQ 500 MATERIAL LIST (8) (continued) PART NAME PART NO. QUANTITY DESCRIPTION COLOR ENHANCEMENT S/W 2165092 1 1 1 1 1 MULTIGESTATION S/W 2165093 1 1 1 1 1 ADVANCED VASCULAR...
  • Page 132 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 12 P9030TA LOGIQ 500 MATERIAL LIST (9) PART NAME PART NO. QUANTITY DESCRIPTION OPERATOR CONSOLE ASSY 2250227 120V, NTSC VER5.00 OPERATOR CONSOLE ASSY 2250232 220V, PAL VER5.00 OPERATOR CONSOLE ASSY 2250228 220V, PAL VER5.00...
  • Page 133 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA LOGIQ 500 MATERIAL LIST (10) PART NAME PART NO. QUANTITY DESCRIPTION OPERATOR CONSOLE ASSY 2275352 120V, NTSC VER6.00 OPERATOR CONSOLE ASSY 2275357 220V, PAL VER6.00 OPERATOR CONSOLE ASSY 2275353 220V, PAL VER6.00...
  • Page 134 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 13 P9030TA OPERATOR CONSOLE ASSY MONITOR BLOCK (FRU100–117) PC BOARDS See ILLUSTRATION 6–2 (FRU601–620) 15–INCH MONITOR BLOCK See ILLUSTRATION 6–9 (FRU170–182) See ILLUSTRATION 6–2b NECK BLOCK KEYBOARD BLOCK (FRU201–203) See ILLUSTRATION 6–4 (FRU400–412)
  • Page 135 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA OPERATOR CONSOLE ASSY 1/9 ILLUSTRATION 6–2 6–18 RENEWAL PARTS...
  • Page 136 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 137 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY 1/9b (15–INCH MONITOR) ILLUSTRATION 6–2b 6–20 RENEWAL PARTS...
  • Page 138: Crt Cable Set (Fru No. 117)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY 2250227, 2250232, 2250228, 2250226, 2250229, 2250230, 2250231 : V5.00 2275352, 2275357, 2275353, 2275351, 2275354, 2275355, 2275356 : V6.00 MATERIAL LIST (1/9b) SECTION PART NAME PART NO. QTY.
  • Page 139 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA OPERATOR CONSOLE ASSY 2/9 ILLUSTRATION 6–3 6–22 RENEWAL PARTS...
  • Page 140 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 141 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA OPERATOR CONSOLE ASSY 3/9 316 302 318A ILLUSTRATION 6–4 6–24 RENEWAL PARTS...
  • Page 142 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 143 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 144 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA This page is left blank intentionally. 6–27 RENEWAL PARTS...
  • Page 145 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 11 P9030TA OPERATOR CONSOLE ASSY 4/9 Note The PCB included in the LED SW KEYBOARD2 SET has no openings for the Gain Encoder because the Gain Encoder is mounted on that PCB.
  • Page 146 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 147 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 148 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–31 RENEWAL PARTS...
  • Page 149 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY 5/9 FRU508 FRU501 FRU501B ILLUSTRATION 6–6 6–32 RENEWAL PARTS...
  • Page 150 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 151 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 152 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–35 RENEWAL PARTS...
  • Page 153 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY 6/9 516 or 582 FRU513 ILLUSTRATION 6–7 6–36 RENEWAL PARTS...
  • Page 154 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 155 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 156 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 157 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 158 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 159 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 160 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 161 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY 7/9 614 Nest Mother Assy 605B ACFP Assy 606 TLMC Assy ACFP2 ASSY ACFP3 ASSY 607 DDSC Assy DDSC2 ASSY DDSC3 Assy 609 VIPB Assy VIPB2 ASSY...
  • Page 162 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 163 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 164 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 165 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 166 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 167 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 168 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 169 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 170 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–53 RENEWAL PARTS...
  • Page 171 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY 8/9 ILLUSTRATION 6–9 6–54 RENEWAL PARTS...
  • Page 172 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 173 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 174 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 175 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 176 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 177 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 178 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 179 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 180 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA OPERATOR CONSOLE ASSY P9522BA, P9522CA, P9522AA, 2113639, 2120175, 2120176 : V1.11 2126615, 2126624, 2126607, 2131676, 2126617, 2126626 : V2.00 2149595, 2149597, 2149593, 2149594, 2149596, 2149588 : V2.20 2164533, 2164534, 2164531, 2164532, 2164535, 2164536 : V3.00...
  • Page 181 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–64 RENEWAL PARTS...
  • Page 182 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2 DISASSEMBLY/RE–ASSEMBLY WARNING! ONLY QUALIFIED SERVICE PERSONNEL SHOULD REMOVE ANY COVERS OR PANELS. ELECTRICAL HAZARDS EXISTS AT SEVERAL POINTS INSIDE. BECOME THOROUGHLY FA- MILIAR WITH ALL HAZARDOUS VOLTAGES AND HIGH CURRENT LEVELS TO AVOID ACCI-...
  • Page 183 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–1 Monitor Assy (FRU No. 100) Time Required 0.25 Hours Tool Required Screwdriver Hexagonal Wrench Procedure WARNING! PERSONAL INJURY HAZARD. VIDEO MONITOR SUPPORT ARM IS SPRING LOADED. RE- LEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVE- MENT.
  • Page 184 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–1 Monitor Assy (FRU No. 100) (continued) Monitor Assy MONITOR BLOCK ASSY DISASSEMBLY ILLUSTRATION 6–10 6–67 RENEWAL PARTS...
  • Page 185 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–2 Monitor Cover Set (FRU No. 101) Time Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–11 on page 6–69. Turn OFF the system. Unscrew six screws ( 1 – 6 ).
  • Page 186 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–2 Monitor Cover Set (FRU No. 101) (continued) Monitor Cover Top (106) Monitor Cover Right (105) Monitor Cover Left (104) Escutcheon Assy (102) Escutcheon Front Door (108) MONITOR COVER SET DISASSEMBLY ILLUSTRATION 6–11...
  • Page 187 Escutcheon Assy, be sure to attach the proper label on the Escutcheon Assy as follows: a. Choose the proper label, “LOGIQ 500” or “LOGIQ 500 MD” indicating the same model name as the label on the removed Escutcheon.
  • Page 188 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–2 Monitor Cover Set (FRU No. 101) (continued) d. Press down the label all around. e. Remove the upper pasteboard from the label. refer to ILLUSTRATION 6–14. REMOVING UPPER PASTEBOARD ILLUSTRATION 6–14...
  • Page 189 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–3 Escutcheon Latch (FRU No. 103) Time Required 10 Minute Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–15. Turn OFF the system. Remove the Monitor Cover Set (FRU 101). Refer to 6–2–2 on page 6–68.
  • Page 190 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–4 Escutcheon Front Door (FRU No. 108) Time Required 1 Minute Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–16. Turn OFF the system. Pull out the Escutcheon Front Door (FRU 108) from the Escutcheon Assy (FRU 102) as shown in ILLUSTRATION 6–16.
  • Page 191 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–5 CRT Cap Set (FRU No. 109) Time Required 1 Minute Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–17. Turn OFF the system. Remove the Escutcheon Front Door (FRU 108). Refer to 6–2–4 on page 6–73.
  • Page 192 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–6 CRT Filter (FRU No. 110) Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–18. Turn OFF the system. Remove the two CRT Filter Clamps ( 1 and 2 : FRU 111).
  • Page 193 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–7 Accessory Assy (FRU No. 112) Time Required 0.5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–19. Turn OFF the system. Remove the Monitor Cover Set (FRU 101). Refer to 6–2–2 on page 6–68.
  • Page 194 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–8 Speaker Assy (FRU No. 113) Time Required 0.5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–20. Turn OFF the system. Remove the Monitor Cover Left (FRU 104) and Monitor Cover Right (FRU 105). Refer to 6–2–2 on page 6–68.
  • Page 195 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–9 CRT Lamp Assy (FRU No. 114) Time Required 0.5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–21. Turn OFF the system. Remove the Monitor Cover Left (FRU 104) and Monitor Cover Right (FRU 105). Refer to 6–2–2 on page 6–68.
  • Page 196 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–79 RENEWAL PARTS...
  • Page 197 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–10 CRT Assy (FRU No. 116) Time Required 0.5 Hours Tool Required Screwdriver Hexagonal Wrench Procedure WARNING! PERSONAL INJURY HAZARD. VIDEO MONITOR SUPPORT ARM IS SPRING LOADED. RE- LEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVE- MENT.
  • Page 198 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–10 CRT Assy (FRU No. 116) (continued) CRT ASSY DISASSEMBLY ILLUSTRATION 6–22 Note The CRT Assy does not include the tilt mechanism. 6–81 RENEWAL PARTS...
  • Page 199 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–11 CRT Cable Set (FRU No. 117) Time Required 0.5 Hour Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–23. Turn OFF the system. Remove the Monitor Assy (FRU 100). Refer to 6–2–1 on page 6–66.
  • Page 200 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–83 RENEWAL PARTS...
  • Page 201 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–12 Monitor15 Assy (FRU No. 170) Time Required 0.25 Hours Tool Required Screwdriver Hexagonal Wrench Procedure WARNING! PERSONAL INJURY HAZARD. VIDEO MONITOR SUPPORT ARM IS SPRING LOADED. RE- LEASING ARM WHEN MONITOR IS NOT INSTALLED WILL CAUSE SUDDEN UPWARD MOVE- MENT.
  • Page 202 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–12 Monitor15 Assy (FRU No. 170) (continued) MONITOR15 BLOCK ASSY DISASSEMBLY ILLUSTRATION 6–24 6–85 RENEWAL PARTS...
  • Page 203: Monitor15 Cover Set (Fru No. 171)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–13 Monitor15 Cover Set (FRU No. 171) Time Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–25 on page 6–87. Turn OFF the system. Unscrew two screws ( 1 – 2 ).
  • Page 204 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–13 Monitor15 Cover Set (FRU No. 171) (continued) MONITOR COVER SET DISASSEMBLY ILLUSTRATION 6–25 6–87 RENEWAL PARTS...
  • Page 205 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–88 RENEWAL PARTS...
  • Page 206: Speaker Assy (Fru No. 178)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–14 Speaker Assy (FRU No. 178) Time Required 0.5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–26. Turn OFF the system. Remove the Monitor Cover L2 Assy (FRU 173) and Monitor Cover R2 Assy (FRU 174). Refer to 6–2–13 on page 6–86.
  • Page 207: Task Lamp (Fru No. 180)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–15 TASK Lamp (FRU No. 180) Time Required 0.5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–27. Turn OFF the system. Unscrew two screws ( 1 and 2 ).
  • Page 208: Crt Cable Set (Fru No. 182)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–16 CRT Cable Set (FRU No. 182) Time Required 0.5 Hour Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–28. Turn OFF the system. Remove the Monitor15 Assy (FRU 170). Refer to 6–2–12 on page 6–84.
  • Page 209: Neck Assy (Fru No. 201)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–17 Neck Assy (FRU No. 201) Time Required 0.6 Hour Tool Required Screwdriver Hex Wrench Procedure Refer to ILLUSTRATION 6–29 on page 6–93. Turn OFF the system and unplug the power cord from outlet.
  • Page 210 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–17 Neck Assy (FRU No. 201) (continued) FG: Frame Ground NECK ASSY DISASSEMBLY ILLUSTRATION 6–29 LINEAR SLIDE COVER DISASSEMBLY ILLUSTRATION 6–30 6–93 RENEWAL PARTS...
  • Page 211: Neck Grip (Fru No. 202)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–18 Neck Grip (FRU No. 202) Time Required 5 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–31. Turn OFF the system. Unscrew two hexagonal screws ( 1 and 2 ).
  • Page 212: Neck Space Plate (Fru No. 203)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–19 Neck Space Plate (FRU No. 203) Time Required 0.25 Hours Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–32. Turn OFF the system. Remove the Monitor Assy (FRU No. 100) from the Neck Assy. Refer to 6–2–1 on page 6–66.
  • Page 213 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–96 RENEWAL PARTS...
  • Page 214: Side Cover Left (Fru No. 301)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–20 Side Cover Left (FRU No. 301) Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–33. Turn OFF the system. Remove four screw caps and unscrew four screws ( 1 – 4 ).
  • Page 215: Side Cover Right (Fru No. 302)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–21 Side Cover Right (FRU No. 302) Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–34. Turn OFF the system. Unscrew four screws and remove the Front Base Cover (FRU No. 307). Refer to 6–2–26 on page 6–103.
  • Page 216: Rear Cover Assy (Fru No. 303)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–22 Rear Cover Assy (FRU No. 303) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–35. Turn OFF the system. Pull out the Air Filter ( 1 ).
  • Page 217: Rear Door Assy (Fru No. 304)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–23 Rear Door Assy (FRU No. 304) Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–36. Turn OFF the system. Open the Rear Door (FRU 304). Unscrew four screws ( 1 – 4 ).
  • Page 218: Rear Door Latch (Fru No. 305)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–24 Rear Door Latch (FRU No. 305) Time Required 4 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–37. Turn OFF the system. Open the Rear Door and unscrew two screws ( 1 and 2 ).
  • Page 219: Rear Door Lock Blk (Fru No. 306)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–25 Rear Door Lock BLK (FRU No. 306) Time Required 4 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–38. Turn OFF the system. Open the Rear Door and unscrew four screws ( 1 – 4 ).
  • Page 220: Front Base Cover (Fru No. 307)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–26 Front Base Cover (FRU No. 307) Time Required 5 Minutes Tool Required Driver Procedure Refer to ILLUSTRATION 6–39. Turn OFF the system. Pull the Air Filter out. Unscrew four screws ( 1 – 4 ).
  • Page 221: Front Cover (Fru No. 308)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–27 Front Cover (FRU No. 308) Time Required 5 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–40. Turn OFF the system. Remove the Front Base Cover (FRU No. 307). Refer to 6–2–26 on page 6–103.
  • Page 222 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–27 Front Cover (FRU No. 308) (continued) Note The Front Cover Assy (2211257) consists of the Front Cover (P9524CE) and the Blind Label. It is necessary to attach the Blind Label on the Front Cover to cover the openings for the HDD LED in case of the console with the new type of 3.5–inch HDD (Quantum...
  • Page 223: Keyboard Bottom Cover (Fru No. 309)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–28 Keyboard Bottom Cover (FRU No. 309) Time Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–42. Turn OFF the system. Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
  • Page 224 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–107 RENEWAL PARTS...
  • Page 225: Top Cover (Fru No. 310)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–29 Top Cover (FRU No. 310) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–43 on page 6–109. Turn OFF the system. Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
  • Page 226 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–29 Top Cover (FRU No. 310) (continued) Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î...
  • Page 227: Front Bumper Set (Fru No. 312)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–30 Front Bumper Set (FRU No. 312) Time Required 10 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–44. Turn OFF the system. Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
  • Page 228: Corner Guard (Fru No. 313)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–31 Corner Guard (FRU No. 313) Time Required 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–45. Turn OFF the system. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
  • Page 229: Probe Holder (Fru No. 314)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–32 Probe Holder (FRU No. 314) Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–46. Turn OFF the system. Pull the latch located at bottom of the probe holder.
  • Page 230: Gel Holder (Fru No. 315)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–33 Gel Holder (FRU No. 315) Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–47. Turn OFF the system. Pull the Gel Holder out as shown in ILLUSTRATION 6–47.
  • Page 231: Handle (Fru No. 317)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–34 Handle (FRU No. 317) Time Required 10 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–48. Turn OFF the system. Remove the Rear Cover Assy (FRU No. 303). Refer to 6–2–22 on page 6–99.
  • Page 232: Air Filter (Fru No. 318A), Pw Filter (Fru No. 318B)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–35 Air Filter (FRU No. 318A), PW Filter (FRU No. 318B) Time Required 3 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–49. Turn OFF the system. Pull out the two Air Filters ( 1 and 2 ) and Remove the one PW Filter ( 3 ) as shown in ILLUSTRATION 6–49.
  • Page 233: Cable Arm Assy (Fru No. 319)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–36 Cable Arm Assy (FRU No. 319) Time Required 1 Minute Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–50. Turn OFF the system. Pull the Cable Arm Assy out as shown in ILLUSTRATION 6–50.
  • Page 234: Cable Arm Knob (Fru No. 320)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–37 Cable Arm Knob (FRU No. 320) Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–51. Turn OFF the system. Turn each Cable Arm Knob counterclockwise and remove it.
  • Page 235: Front Caster (Fru No. 321)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–38 Front Caster (FRU No. 321) Time Required 15 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–52. Turn OFF the system. Put the system on a box.
  • Page 236: Rear Caster (Fru No. 322)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–39 Rear Caster (FRU No. 322) Time Required 15 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–53. Turn OFF the system. Put the system on a box.
  • Page 237: Caster Cover Set (Fru No. 323)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–40 Caster Cover Set (FRU No. 323) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–54. Turn OFF the system. Unscrew five screws ( 1 – 5 ).
  • Page 238: Keyboard Panel Assy And Keyboard Panel2 Assy (Fru No. 400)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–41 Keyboard Panel Assy and Keyboard Panel2 Assy (FRU No. 400) Time Required 8 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–55. Turn OFF the system. Unscrew three screws ( 1 – 3 ) of the Keyboard Panel Assy or Keyboard Panel2 Assy.
  • Page 239: Keyboard Knob Set (Fru No. 401)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–42 Keyboard Knob Set (FRU No. 401) Time Required 15 Minutes Tool Required Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–56. Turn OFF the system. Unscrew two hexagonal bolts for each encoder knob and remove the seven Encoder Knobs ( 1 – 7 ).
  • Page 240: Trackball (Fru No. 402)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–43 Trackball (FRU No. 402) Time Required 11 Minute Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–57. Turn OFF the system. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
  • Page 241: Tgc Assy (Fru No. 403)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–44 TGC Assy (FRU No. 403) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–58. Turn OFF the system. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
  • Page 242: I/F Board Assy (Fru No. 405)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–45 I/F Board Assy (FRU No. 405) Time Required 25 Minutes Tool Required Screwdriver Hexagonal Socket Wrench Procedure Refer to ILLUSTRATION 6–59. Turn OFF the system. Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
  • Page 243: A/N Keyboard Assy (Fru No. 406)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–46 A/N Keyboard Assy (FRU No. 406) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Socket Wrench Procedure Refer to ILLUSTRATION 6–60. Turn OFF the system. Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
  • Page 244 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–127 RENEWAL PARTS...
  • Page 245: Led Sw Keyboard Set (Fru No. 407)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–47 LED SW Keyboard Set (FRU No. 407) Time Required 40 Minutes Tool Required Screwdriver Hexagonal Socket Wrench Procedure Refer to ILLUSTRATION 6–61 on page 6–129. Turn OFF the system.
  • Page 246 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–47 LED SW Keyboard Set (FRU No. 407) (continued) LED SW KEYBOARD SET DISASSEMBLY ILLUSTRATION 6–61 6–129 RENEWAL PARTS...
  • Page 247: Gain Encoder (Fru No. 408)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–48 Gain Encoder (FRU No. 408) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–62. Turn OFF the system. Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
  • Page 248: Vfd Assy (Fru No. 409)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–49 VFD Assy (FRU No. 409) Time Required 20 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–63. Turn OFF the system. Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
  • Page 249: Keyboard Case (Fru No. 410)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–50 Keyboard Case (FRU No. 410) Time Required 1 Hour Tool Required Screwdriver Hexagonal Socket Wrench Procedure Refer to ILLUSTRATION 6–64. Turn OFF the system. Remove the Keyboard Knobs. Refer to 6–2–42 on page 6–122.
  • Page 250: A/N Key-Top Set (Fru No. 411)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–51 A/N Key–top Set (FRU No. 411) Time Required 5 Minutes Tool Required Tweezers Procedure Refer to ILLUSTRATION 6–65. Turn OFF the system. Insert the tweezers into the bottom of key–top.
  • Page 251 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–51 A/N Key–top Set (FRU No. 411) (continued) Key–top for English, French, German, Spanish, Italian, Portuguese, Swedish and Danish Key–top for Greek ILLUSTRATION 6–66 6–134 RENEWAL PARTS...
  • Page 252 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–51 A/N Key–top Set (FRU No. 411) (continued) Key–top for Turkish Key–top for Russian ILLUSTRATION 6–66 (continued) 6–135 RENEWAL PARTS...
  • Page 253: Led Sw Key-Top Set (Fru No. 412)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–52 LED SW Key–top Set (FRU No. 412) Time Required 40 Minutes Tool Required Screwdriver Hexagonal Socket Wrench Procedure Refer to ILLUSTRATION 6–67. Turn OFF the system. Remove the Keyboard Panel Assy. Refer to 6–2–41 on page 6–121.
  • Page 254 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–137 RENEWAL PARTS...
  • Page 255: Rear Conn Panel Assy (Fru No. 501) And Rear Conn Panel2 Assy (Fru No. 501B)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–53 Rear CONN Panel Assy (FRU No. 501) and Rear CONN Panel2 Assy (FRU No. 501B) Time Required 0.5 Hour Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–68 and ILLUSTRATION 6–69.
  • Page 256: Rear 1 Assy (Fru No. 502)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–53 Rear CONN Panel Assy (FRU No. 501) and Rear CONN Panel2 Assy (FRU No. 501B) (continued) Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
  • Page 257 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–54 Rear 1 Assy (FRU No. 502) Time Required 1 Hour Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–70 on page 6–141. Turn OFF the system. Remove the Rear CONN Panel Assy (FRU 501). Refer to 6–2–53 on page 6–138.
  • Page 258 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–54 Rear 1 Assy (FRU No. 502) (continued) Î Î Î Î Î Î Î Î Î Î Î Î Î Î Î Back of Rear Panel Assy Î Î Î Î Î...
  • Page 259: Rear 2 Assy (Fru No. 503)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–55 Rear 2 Assy (FRU No. 503) Time Required 1 Hours Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–71 on page 6–143. Turn OFF the system. Remove the Rear CONN Panel Assy (FRU 501) or the Rear CONN Panel2 Assy (FRU 501B). Refer to 6–2–53 on page 6–138.
  • Page 260 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–55 Rear 2 Assy (FRU No. 503) (continued) Rear CONN Panel Assy (Rear Side View) Rear CONN Panel Assy Î Î Î Î Î Î (Front Side View) Î Î Î Î...
  • Page 261: Rear 3 Assy (Fru No. 590)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–56 Rear 3 Assy (FRU No. 590) Time Required 1 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–72. Turn OFF the system. Remove the Rear CONN Panel2 Assy (FRU 501B). Refer to 6–2–53 on page 6–138.
  • Page 262: Rear Connector Set (Fru No. 504)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–57 Rear Connector Set (FRU No. 504) Time Required 10 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–73. Turn OFF the system. Remove the Rear CONN Panel Assy (FRU 501). Refer to 6–2–53 on page 6–138.
  • Page 263: Scsi Terminator (Fru No. 505)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–58 SCSI Terminator (FRU No. 505) Time Required 1 Minutes Tool Required Not necessary Procedure Refer to ILLUSTRATION 6–74. Turn OFF the system. Open the Rear Door. Remove the SCSI Terminator.
  • Page 264: Circuit Protector (Fru No. 506: 7A, Fru No. 507: 3A)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–59 Circuit Protector (FRU No. 506: 7A, FRU No. 507: 3A) Time Required 30 Minutes Tool Required Screwdriver Soldering Iron Procedure Refer to ILLUSTRATION 6–75. Turn OFF the system. Remove the Rear CONN Panel Assy (FRU 501) or the Rear CONN Panel2 Assy (FRU 501B). Refer to 6–2–53 on page 6–138.
  • Page 265: Hv Assy (Fru No. 508)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–60 HV Assy (FRU No. 508) Time Required 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–76. Turn OFF the system. Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
  • Page 266: Hv Fan Set (Fru No. 509)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–61 HV Fan Set (FRU No. 509) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–77. Turn OFF the system. Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
  • Page 267: Ecg Board Assy (Fru No. 510)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–62 ECG Board Assy (FRU No. 510) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–78. NOTE The new type of ECG Board Assy is installed from production unit of LOGIQ500 console with soft- ware version 5.00 y or later.
  • Page 268 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–62 ECG Board Assy (FRU No. 510) (continued) Note New type of ECG Board Assy (2207806) has the reserved connector (P69). This connector is not used for the LOGIQ 500 console.
  • Page 269: Ecg Panel Assy (Fru No. 511)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–63 ECG Panel Assy (FRU No. 511) Time Required 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–80. NOTE The new type of ECG Panel Assy is installed from production unit of LOGIQ500 console with soft- ware Version 5.00 y or later.
  • Page 270 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–63 ECG Panel Assy (FRU No. 511) (continued) NOTE To remove the new type of ECG Panel Assy, unscrew two screws ( ). Loosen two screws ). Remove the new type of ECG Panel Assy after disconnect three connectors ( –...
  • Page 271: Nest Fan Assy (Fru No. 512)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–64 Nest Fan Assy (FRU No. 512) Time Required 45 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–83. Turn OFF the system. Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
  • Page 272 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–64 Nest Fan Assy (FRU No. 512) (continued) Disconnect two connectors ( 1 and 2 ). Unscrew three screws ( 3 – 5 ). Unscrew three screws and remove the frame plate ( 6 ).
  • Page 273: Probe Conn Set Assy (Fru No. 513)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–65 Probe CONN Set Assy (FRU No. 513) Time Required 40 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–84 and ILLUSTRATION 6–85. Turn OFF the system. Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
  • Page 274 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–65 Probe CONN Set Assy (FRU No. 513) (continued) 14. Remove the Blind Metal Plate. Note When the customer uses the P509 probe, the Probe CONN 1 Assy is connected to the RCTL2 Assy with the cable.
  • Page 275: Shield Finger Long (Fru No. 514), Shield Finger Short (Fru No. 515)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–66 Shield Finger Long (FRU No. 514), Shield Finger Short (FRU No. 515) Time Required 45 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–86. Turn OFF the system. Remove the Probe CONN Set Assy (FRU 513). Refer to 6–2–65 on page 6–156.
  • Page 276: Probe Conn 1 Assy (Fru No. 516)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–67 Probe CONN 1 Assy (FRU No. 516) Time Required 45 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–87. Turn OFF the system. Remove the Probe CONN Set Assy (FRU 513). Refer to 6–2–65 on page 6–156.
  • Page 277: Probe Conn 2 Assy (Fru No. 517)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–68 Probe CONN 2 Assy (FRU No. 517) Time Required 45 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–88. Turn OFF the system. Remove the Probe CONN Set Assy (FRU 513). Refer to 6–2–65 on page 6–156.
  • Page 278: Fdd (Fru No. 518), Fdd Holder Assy (Fru No. 519)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–69 FDD (FRU No. 518), FDD Holder Assy (FRU No. 519) Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–89. Turn OFF the system. Remove the Keyboard Panel Assy (FRU 400). Refer to 6–2–41 on page 6–121.
  • Page 279: Hdd Assy (Fru No. 520), Hdd Holder Assy (Fru No. 521)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521) Time Required 1.5 Hours Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–90 on page 6–163. Make sure the customer data is saved on disks (floppy disks for system software version 1.20 or former console and MO disk for 2.00 or later console).
  • Page 280 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521) (continued) HDD ASSY AND HDD HOLDER ASSY DISASSEMBLY ILLUSTRATION 6–90 6–163 RENEWAL PARTS...
  • Page 281 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521) (continued) 3.5–Inch HDD Disassembly Refer to ILLUSTRATION 6–91 on page 6–165. Make sure the customer data is saved on MO disk. Refer to 4–3–3 (D) on page 4–21, USER DATA BACK UP.
  • Page 282 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521) (continued) 3.5–Inch HDD Disassembly (continued) HDD ASSY AND HDD HOLDER2 ASSY DISASSEMBLY ILLUSTRATION 6–91 6–165 RENEWAL PARTS...
  • Page 283 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521) (continued) 3.5–Inch HDD Disassembly (continued) Note Before installing new 3.5–inch HDD Assy, make sure that the jumper switches located at the bottom of new 3.5–inch HDD are set properly and the terminator module IC is NOT mounted on RP1 as shown...
  • Page 284 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–70 HDD Assy (FRU No. 520), HDD Holder Assy (FRU No. 521) (continued) 3.5–Inch HDD Disassembly (continued) Note The setting of jumper switches for the new 3.5–inch HDD Assy (Quantum Fireball EL) is different from others.
  • Page 285: Hddb Assy (Fru No. 522)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–71 HDDB Assy (FRU No. 522) Note The HDDB Assy is not used for the console with the 3.5–inch HDD Assy. Time Required 15 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–95.
  • Page 286: Hdd Led Assy (Fru No. 523)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–72 HDD LED Assy (FRU No. 523) Note The HDD LED Assy is not available on the console with the new type of 3.5–inch HDD Assy (Quan- tum FireBall 1280S or newer ).
  • Page 287: Power Switch Assy (Fru No. 525)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–73 Power Switch Assy (FRU No. 525) Time Required 10 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–97. NOTE The new type of Power Switch Assy is installed from production unit of LOGIQ500 console with soft- ware version 5.00 y or later.
  • Page 288 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–73 Power Switch Assy (FRU No. 525) NOTE To remove the new type of Power Switch Assy, disconnect one connector ( 1 ). Unscrew the nine screws ( 2 –...
  • Page 289: Modd (Fru No. 540), Modd Fan (Fru No. 541)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–74 MODD (FRU No. 540), MODD Fan (FRU No. 541), MODD Holder Assy (FRU No. 542) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–99 on page 6–173.
  • Page 290 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–74 MODD (FRU No. 540), MODD Fan Assy (FRU No. 541), MODD Holder Assy (FRU No. 542) (continued) MODD, MODD FAN ASSY, AND MODD HOLDER ASSY REMOVAL ILLUSTRATION 6–99 6–173...
  • Page 291 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–74 MODD (FRU No. 540), MODD Fan Assy (FRU No. 541), MODD Holder Assy (FRU No. 542) (continued) Note Make sure that the DIP switches and jumper switches located on the MODD are set as follows before you install a new MODD onto the system.
  • Page 292 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–74 MODD (FRU No. 540), MODD Fan Assy (FRU No. 541), MODD Holder Assy (FRU No. 542) (continued) Note If the 3.5–inch HDD is installed on the console, set the jumper switches as shown in ILLUSTRATION 6–102.
  • Page 293 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–74 MODD (FRU No. 540), MODD Fan Assy (FRU No. 541), MODD Holder Assy (FRU No. 542) (continued) Note If the 3.5–inch HDD is installed on the console, set the jumper switches as shown in ILLUSTRATION 6–103.
  • Page 294 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–74 MODD (FRU No. 540), MODD Fan Assy (FRU No. 541), MODD Holder Assy (FRU No. 542) (continued) Emergency Ejection When your disk is not mounted and the eject button does not function, eject the disk following the procedure below: CAUTION Use the emergency ejection procedure only in emergency situations.
  • Page 295 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–178 RENEWAL PARTS...
  • Page 296 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–179 RENEWAL PARTS...
  • Page 297: Board(S) (Fru No. 601 Through 610)

    SYSTEM ID ENTRY/DISPLAY in Chapter 4 of this manual. For the procedure of installing system ID number into the system software version 1.20 or former con- sole, refer to LOGIQ 500 Proprietary Manual, P9030TB. Note If the customer uses the P509 probe, you have to disconnect the connector from the RCTL2 Assy before removing the RCTL2 Assy.
  • Page 298 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–75 P.C. Boards (FRU No. 601 through 610) (continued) P.C. BOARD(S) REMOVAL ILLUSTRATION 6–105 6–181 RENEWAL PARTS...
  • Page 299: Time Keeper Ram (Fru No. 611)

    After replacing a time keeper RAM with new one, it is necessary to install the system ID number into new time keeper RAM and set up two items, Video Signal and Power Frequency among four system preset parameters. For set-up procedures of system preset parameter, refer to LOGIQ 500 Propri- etary Manual, P9030TB.
  • Page 300 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–76 Time Keeper RAM (FRU No. 611) (continued) TIME KEEPER RAM REMOVAL ILLUSTRATION 6–106 6–183 RENEWAL PARTS...
  • Page 301: Time Keeper Battery (Fru No. 612)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–77 Time Keeper Battery (FRU No. 612) Time Required 1 Hour Tool Required Screwdriver IC Remover Procedure Refer to ILLUSTRATION 6–107 on page 6–185. Turn OFF the system. Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
  • Page 302 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–77 Time Keeper Battery (FRU No. 612) (continued) TIME KEEPER BATTERY REMOVAL ILLUSTRATION 6–107 6–185 RENEWAL PARTS...
  • Page 303: Time Keeper Ram (For Mvme167-002B) (Fru No. 616)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–78 Time Keeper RAM (for MVME167–002B) (FRU No. 616) Time Required 1 Hour Tool Required Screwdriver IC Remover Procedure Refer to ILLUSTRATION 6–108 on page 6–187. Turn OFF the system.
  • Page 304 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–78 Time Keeper RAM (for MVME167–002B) (FRU No. 616) (continued) TIME KEEPER RAM (for MVME167–002B) REMOVAL ILLUSTRATION 6–108 6–187 RENEWAL PARTS...
  • Page 305: Omem Assy (Fru No. 613)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–79 OMEM Assy (FRU No. 613) Time Required 20 Minutes Tool Required Screwdriver Procedure CAUTION Do not bend or flex the boards when mounting/dismounting the piggyback boards. Surface mount IC boards are very susceptible to damage from flex/torque.
  • Page 306 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–79 OMEM Assy (FRU No. 613) (continued) OMEM ASSY DISASSEMBLY ILLUSTRATION 6–109 6–189 RENEWAL PARTS...
  • Page 307: Nest Mother Assy (Fru No. 614)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–80 Nest Mother Assy (FRU No. 614) Time Required 2 Hours Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–110 on page 6–192. CAUTION An electrostatic discharge may damage a component. Turn OFF power and wear the wrist strap before you remove a circuit boards.
  • Page 308 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–80 Nest Mother Assy (FRU No. 614) (continued) 11. Remove the Rear CONN Panel Assy (FRU 501). Refer to 6–2–53 on page 6–138. 12. Remove the HDD LED Assy (FRU 523). Refer to 6–2–72 on page 6–169.
  • Page 309 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–80 Nest Mother Assy (FRU No. 614) (continued) NEST MOTHER ASSY DISASSEMBLY ILLUSTRATION 6–110 6–192 RENEWAL PARTS...
  • Page 310: Mother Ic (Fru No. 615)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–81 Mother IC (FRU No. 615) Time Required 1 Hour Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–111. Turn OFF the system. Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
  • Page 311: Dnid2 Assy (Fru No. 620)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–82 DNID2 Assy (FRU No. 620) Note The DNID2 Assy is supplied with the optional Single CWD upgrade kit. Time Required 20 Minutes Tool Required Screwdriver Procedure CAUTION Do not bend or flex the boards when mounting/dismounting the piggyback boards. Surface mount IC boards are very susceptible to damage from flex/torque.
  • Page 312 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–82 DNID2 Assy (FRU No. 620) (continued) DNID2 ASSY DISASSEMBLY ILLUSTRATION 6–112 6–195 RENEWAL PARTS...
  • Page 313 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–196 RENEWAL PARTS...
  • Page 314 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–197 RENEWAL PARTS...
  • Page 315: A1 Ps +5V 32A (Fru No. 701)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–83 A1 PS +5V 32A (FRU No. 701) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–113 on page 6–199. Turn OFF the system.
  • Page 316 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–83 A1 PS +5V 32A (FRU No. 701) (continued) A1 PS +5V 32A DISASSEMBLY ILLUSTRATION 6–113 6–199 RENEWAL PARTS...
  • Page 317: A2 Ps -5.2V 11A (Fru No. 702)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–84 A2 PS –5.2V 11A (FRU No. 702) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–114 on page 6–201. Turn OFF the system.
  • Page 318 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–84 A2 PS –5.2V 11A (FRU No. 702) (continued) A2 PS –5.2V 11A DISASSEMBLY ILLUSTRATION 6–114 6–201 RENEWAL PARTS...
  • Page 319: A3 Ps +12V 8.4A (Fru No. 703)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–85 A3 PS +12V 8.4A (FRU No. 703) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–115 on page 6–203. Turn OFF the system.
  • Page 320 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–85 A3 PS +12V 8.4A (FRU No. 703) (continued) A3 PS +12V 8.4A DISASSEMBLY ILLUSTRATION 6–115 6–203 RENEWAL PARTS...
  • Page 321: A4 Ps -12V 2.5A (Fru No. 704)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–86 A4 PS –12V 2.5A (FRU No. 704) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–116 on page 6–205. Turn OFF the system.
  • Page 322 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–86 A4 PS –12V 2.5A (FRU No. 704) (continued) A4 PS –12V 2.5A DISASSEMBLY ILLUSTRATION 6–116 6–205 RENEWAL PARTS...
  • Page 323: A5 Ps +5V 30A (Fru No. 705)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–87 A5 PS +5V 30A (FRU No. 705) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–117 on page 6–207. Turn OFF the system.
  • Page 324 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–87 A5 PS +5V 30A (FRU No. 705) (continued) A5 PS +5V 30A DISASSEMBLY ILLUSTRATION 6–117 6–207 RENEWAL PARTS...
  • Page 325: A7 Sw Cont Assy (Fru No. 706)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–88 A7 SW Cont Assy (FRU No. 706) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–118 on page 6–209. Turn OFF the system.
  • Page 326 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–88 A7 SW Cont Assy (FRU No. 706) (continued) A7 SW CONT ASSY DISASSEMBLY ILLUSTRATION 6–118 6–209 RENEWAL PARTS...
  • Page 327: A8 Pfc Assy (Fru No. 707)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–89 A8 PFC Assy (FRU No. 707) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–119 on page 6–211. Turn OFF the system. Remove the Front Base Cover (FRU 307). Refer to 6–2–26 on page 6–103.
  • Page 328 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–89 A8 PFC Assy (FRU No. 707) (continued) A8 PFC ASSY DISASSEMBLY ILLUSTRATION 6–119 6–211 RENEWAL PARTS...
  • Page 329: A10 Hv Reg Assy (Fru No. 708)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–90 A10 HV REG Assy (FRU No. 708) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench (Key Size: 3 mm) Procedure Refer to ILLUSTRATION 6–120 on page 6–213. Turn OFF the system.
  • Page 330 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–90 A10 HV REG Assy (FRU No. 708) (continued) A10 HV REG ASSY DISASSEMBLY ILLUSTRATION 6–120 6–213 RENEWAL PARTS...
  • Page 331: Rl1 & 2 Assy (Fru No. 709)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–91 RL1 & 2 Assy (FRU No. 709) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–121 on page 6–215. Turn OFF the system. Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
  • Page 332 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–91 RL1 & 2 Assy (FRU No. 709) (continued) RL1 & 2 ASSY DISASSEMBLY ILLUSTRATION 6–121 6–215 RENEWAL PARTS...
  • Page 333: Rl3 (Fru No. 710)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–92 RL3 (FRU No. 710) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–122 on page 6–217. Turn OFF the system. Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
  • Page 334 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–92 RL3 (FRU No. 710) (continued) RL3 DISASSEMBLY ILLUSTRATION 6–122 6–217 RENEWAL PARTS...
  • Page 335: Ps Fan Set (Fru No. 713)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–93 PS Fan Set (FRU No. 713) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–123 on page 6–219. Turn OFF the system. Remove the Side Cover Left (FRU 301). Refer to 6–2–20 on page 6–97.
  • Page 336 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–93 PS Fan Set (FRU No. 713) (continued) PS FAN SET DISASSEMBLY ILLUSTRATION 6–123 6–219 RENEWAL PARTS...
  • Page 337: Breaker 20A (Fru No. 714), Breaker 7.5A (Fru No. 715)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–94 Breaker 20A (FRU No. 714), Breaker 7.5A (FRU No. 715) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–124 on page 6–221. Turn OFF the system.
  • Page 338 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–94 Breaker 20A (FRU No. 714), Breaker 7.5A (FRU No. 715) (continued) BREAKER DISASSEMBLY ILLUSTRATION 6–124 6–221 RENEWAL PARTS...
  • Page 339: Inlet Unit 120V, Inlet Unit 220V (Fru No. 720)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–95 Inlet Unit 120V, Inlet Unit 220V (FRU No. 720) Time Required 30 Minutes Tool Required Screwdriver Procedure Refer to ILLUSTRATION 6–125 on page 6–223. Turn OFF the system. Remove the Rear Cover Assy (FRU 303). Refer to 6–2–22 on page 6–99.
  • Page 340 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–95 Inlet Unit 120V, Inlet Unit 220V (FRU No. 720) (continued) INLET UNIT DISASSEMBLY ILLUSTRATION 6–125 6–223 RENEWAL PARTS...
  • Page 341: Inrush Limiter 120V, Inrush Limiter 220V (Fru No. 721)

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–96 Inrush Limiter 120V, Inrush Limiter 220V (FRU No. 721) Time Required 30 Minutes Tool Required Screwdriver Hexagonal Wrench Procedure Refer to ILLUSTRATION 6–126 on page 6–225. Turn OFF the system.
  • Page 342 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–2–96 Inrush Limiter 120V, Inrush Limiter 220V (FRU No. 721) (continued) INTUSH LIMITER DISASSEMBLY ILLUSTRATION 6–126 6–225 RENEWAL PARTS...
  • Page 343 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–226 RENEWAL PARTS...
  • Page 344: Fuse Replacement

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3 FUSE REPLACEMENT 6–3–1 Introduction LOGIQ 500 has sixteen replaceable fuses. The table below states the items of those fuses. TABLE 6–1 REPLACEABLE FUSES LOCATION Ref. No. RATING REMARKS REAR1 ASSY...
  • Page 345 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3–2 Replacement Procedures (continued) (A) F1 on REAR1 Assy 1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902). 2. Remove the Side Cover Right (FRU302). Refer to sub–section 6–2–21 on page 6–98.
  • Page 346 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3–2 Replacement Procedures (continued) (B) F1 and F2 located on TB3 1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902). 2. Remove the PS Cover Left. Refer to sub–section 6–2–90 on page 6–212.
  • Page 347 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3–2 Replacement Procedures (continued) (C) F3 in PS Unit Note This F3 fuse is used only on the 120V unit. 1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902).
  • Page 348 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3–2 Replacement Procedures (continued) (D) F1 and F2 on SW Cont Assy 1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902). 2. Remove the PS Cover Right. Refer to sub–section 6–2–83 on page 6–198.
  • Page 349 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3–2 Replacement Procedures (continued) (E) F1 on PFC Assy 1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902). 2. Remove the PS Cover Right. Refer to sub–section 6–2–83 on page 6–198.
  • Page 350 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–3–2 Replacement Procedures (continued) (F) F1 through F9 on HV REG Assy 1. Turn the system power OFF. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902). 2. Remove the PS Cover Left. Refer to sub–section 6–2–90 on page 6–212.
  • Page 351 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–234 RENEWAL PARTS...
  • Page 352: Introduction

    6–4–1 Introduction This section describes the procedures for installing the system software into the hard disk on LOGIQ 500. Note You can upgrade the version of LOGIQ 500 system software up using the Software Version Up func- tion of LOGIQ 500 Service Software.
  • Page 353: Preparing Before Installation

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–4–2 Preparing before Installation Turn the system power OFF. Unplug the system. Follow the OSHA Lockout/Tagout Requirements (29 CFR 1910.147, Direction 46–015902). Replace the hard disk with new one if necessary. Refer to sub-section 6–2–70 on page 6–162 for the replacement procedure of hard disk.
  • Page 354 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–4–2 Preparing before Installation (continued) Power the Laptop PC ON. Start the terminal emulation software. 10. Make sure that the communication parameters are set as follows: Baud Rate 9600 bps...
  • Page 355: Initializing Hard Disk

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–4–3 Initializing Hard Disk (continued) Type the following command and press the return key. VTSH>mediainit_0.0_PHILE_240000000_20_10000_4000 Note For the console with software version 4.10 or later, use the following command: VTSH>mediainit_0.0_PHILE_540000000_20_10000_4000 Note The character “_”...
  • Page 356: Installing Software

    3–1 Set the Country/Zone properly. It takes approximately 40 minutes to complete this step. 3–2. Set other three parameters, Video Signal, Power Frequency and Modem Kind. 3–3. Reset the system using System Reset function of the Service Menu. Refer to LOGIQ 500 Proprietary Manu- al, P9030TB.
  • Page 357: Installing Using Mo Disk

    Insert the system software MO disk into the MO disk drive of LOGIQ 500 console. Turn the system power ON. LOGIQ 500 system does not perform the ordinary power-up sequence and the message “Now starting the ver- sion up process!!” is displayed on the CRT monitor screen.
  • Page 358 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–241 RENEWAL PARTS...
  • Page 359 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA This page is left blank intentionally. 6–242 RENEWAL PARTS...
  • Page 360: Power Supply Replacement

    P9030TA 6–5 POWER SUPPLY REPLACEMENT 6–5–1 Introduction This section describes the procedures for replacing the power supply unit of LOGIQ 500. 6–5–2 Replacement Procedure READ and UNDERSTAND these instructions thoroughly before proceeding with the replacement. Perform each step in sequence and check it off when completed. If a problem occurs after completion, go back through the procedure and check for implementation errors before contacting your Ultrasound Region Engineer.
  • Page 361 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) (A) Removing Monitor, Covers, and Neck Assy Turn the system power OFF and unplug the unit. Lift and set the Monitor to the highest position by pushing down the Up/down Release Button located on the Neck Assy.
  • Page 362 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) (B) Disconnecting Cables Disconnect fourteen cables listed below. Cut the tie wraps off if necessary. For each cable, refer to ILLUSTRATION 6–137. a. Unscrew eight hexagonal bolts and remove eight cables (two CB71, two CB73, two CB75, and two CB77) from the Mother Assy.
  • Page 363 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) (C) Replacing Power Supply Place three jacks at the positions indicated in ILLUSTRATION 6–138. Make sure that the jacks are evenly seated on the floor and the flat surface of the console.
  • Page 364 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) Lift the system up using the jacks by approximately 30mm. Be careful to keep the console horizontally as much as you can. Insert the pallet between the floor and the Power Supply unit. Refer to ILLUSTRATION 6–139.
  • Page 365 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) Unscrew six hexagonal bolts. Three bolts are located at the right side and other three bolts are located at the left side. Lift the console up using the jacks by approximately 100mm to pull the cables out of the console. Refer to ILLUSTRATION 6–140.
  • Page 366 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) Slide the pallet sideways. POWER SUPPLY UNIT REMOVAL (b) ILLUSTRATION 6–141 Remove the power supply unit and place new one on the pallet. WARNING! PERSONAL INJURY HAZARD! POWER SUPPLY UNIT WEIGHS APPROXIMATELY 51 KILO- GRAMS (112 POUNDS).
  • Page 367 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–5–2 Replacement Procedure (continued) (D) Final Procedure Connect fourteen cables with the appropriate connectors. For the details of cables, refer to 6–5–2 (B), Discon- necting Cables on page 6–219. Re–assemble the removal parts. Refer to section 6–2, DISASSEMBLY/RE-ASSEMBLY.
  • Page 368 REV 14 P9030TA 6–6 REPLACING 2.5–INCH HDD ASSY WITH 3.5–INCH HDD ASSY 6–6–1 Introduction This section provides the installation procedures of the 3.5–Inch HDD Assy using the Installation Kit for LOGIQ 500 console. 6–6–2 Time Required 1 hour 6–6–3 Parts Required...
  • Page 369 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables (continued) Cut the three tie wraps and remove them. Refer to ILLUSTRATION 6–142. REMOVING HDDB ASSY ILLUSTRATION 6–142 6–252...
  • Page 370 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables (continued) Unscrew four screws and remove the MODD Cover Metal. REMOVING MODD COVER METAL ILLUSTRATION 6–143 6–253 RENEWAL PARTS...
  • Page 371 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables (continued) 10. Disconnect the CB24M cable. Refer to ILLUSTRATION 6–144. DISCONNECTING CB24 CABLE ILLUSTRATION 6–144 11. Disconnect three cables, CB24M, CB5, and CB46, from the HDDB Assy. Refer to ILLUSTRATION 6–145.
  • Page 372 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (A) Removing 2.5–inch HDD Assy, HDDB Assy and Cables (continued) 13. Unscrew two screws and remove the 2.5–inch HDD Assy with HDD Holder Assy. Refer to ILLUSTRATION 6–146.
  • Page 373 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (B) Connecting Cables Note The Cable Bracket is installed from production unit of LOGIQ500 console with software version 5.00 y or later. Insert and thread the CB24N Cable into the slit of the Cable Bracket as shown in ILLUSTRATION 6–148.
  • Page 374 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (B) Connecting Cables (continued) 17. Connect the CB24N Cable with the P24A connector on the Mother Assy. Refer to ILLUSTRATION 6–149. 18. Connect the CB5N Cable with the P5A connector on the Mother Assy. Refer to ILLUSTRATION 6–149.
  • Page 375 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy 20. Make sure that the four jumper switch are set as shown in ILLUSTRATION 6–150. 21. Make sure that the Terminator Module IC is not mounted on the 3.5–inch HDD Assy. Refer to ILLUSTRATION 6–150.
  • Page 376 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) Note The setting of jumper switches for the new 3.5–inch HDD Assy (Quantum Fireball EL) is different from others. Refer to ILLUSTRATION 6–152 to check and set the jumper switches properly.
  • Page 377 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) 22. Connect the CB45N cable with the 3.5–inch HDD Assy. Refer to ILLUSTRATION 6–153. 23. Connect another side of the CB45N cable to the CB46 cable.
  • Page 378 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) 26. Screw two screws supplied with the HDD Change Kit to install the 3.5–inch HDD Assy with HDD Holder2 Assy on the console.
  • Page 379 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) Note The folding method of CB24N cable is different depending on the type of MODD (U0052ZA or 2187012). Before connecting the CB24N cable, fold the CB24N cable suitably. Refer to ILLUSTRATION 6–155.
  • Page 380 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) 28. Change the settings of jumper switches on the MODD Assy as shown in ILLUSTRATION 6–156. Note The new MODD (2187012) is different in appearance from the old one (U0052ZA).
  • Page 381 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) Note If the 3.5–inch HDD is installed on the console, set the jumper switches as shown in ILLUSTRATION 6–157. SETTINGS OF JUMPER SWITCHES ON MODD (2187012–2) ILLUSTRATION 6–157...
  • Page 382 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) Note If the 3.5–inch HDD is installed on the console, set the jumper switches as shown in ILLUSTRATION 6–157. SETTINGS OF JUMPER SWITCHES ON MODD (2271309) ILLUSTRATION 6–158...
  • Page 383 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 14 P9030TA 6–6–4 Procedures (continued) (C) Installing 3.5–Inch HDD Assy (continued) 31. Bind the CB24N cable with two tie wraps as shown in ILLUSTRATION 6–159. Note Be careful not to bind the CB24N tightly.
  • Page 384: Introduction

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA CHAPTER 7 – PERIODIC MAINTENANCE TABLE OF CONTENTS SECTION TITLE PAGE 7–1 INTRODUCTION ..............
  • Page 385: Periodic Maintenance

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 4 P9030TA This page is left blank intentionally. 7–2 PERIODIC MAINTENANCE...
  • Page 386: Periodic Maintenance Procedure

    Check that all plug–in boards are fully seated in cardcage connector. h. Check operation of the cooling fans and that the airflow paths are clear. LOGIQ 500 has five Fans : three for PC Boards, two for Power Supplies.
  • Page 387 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 9 P9030TA 7–2–2 Cleaning a. Clean the CRT filter. a–1. Remove the CRT filter from the monitor. Refer to section 6–2–6 in Chapter 6, RENEWAL PARTS. CAUTION Equipment damage possibility. Do not use a hydrocarbon glass cleaner on a monitor with an anti-glare shield.
  • Page 388 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 9 P9030TA 7–2–2 Cleaning (Continued) c. Clean two Air Filters ( 1 and 2 ) and one PW Filter ( 3 ) by using a vacuum cleaner and bristled brush. If these filters are damaged, replace them. For the positions of three filters, refer to ILLUSTRATION 7–2.
  • Page 389: User Data Backup

    REV 9 P9030TA 7–2–4 User Data Backup a. Keep user preset data into some floppy disks or MO disk using the User Data Backup of LOGIQ 500 Utility menu. Refer to 4–3–3, Utility Menu in Chapter 4, FUNCTIONAL CHECKS. 7–2–5 Note Visual inspection of probes (“head”, cable and connector)
  • Page 390: Electrical Safety Tests

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3 ELECTRICAL SAFETY TESTS The following tests are performed at the factory and should be performed at the site. These tests are: grounding conti- nuity, chassis leakage current, probe leakage current, and ECG leakage current. All measurements are made with an Electrical Safety Analyzer Model Dale 600/600E built by Dale Technology Corporation or equivalent device.
  • Page 391: Grounding Continuity

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–2 Grounding Continuity CAUTION Electric Shock Hazard. The patient must not be contacted to the equipment during this test. Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case (ILLUSTRATION 7–4).
  • Page 392: Chassis Leakage Current Test

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–3 Chassis Leakage Current Test Definition Leakage current is the electrical current that could flow through the patient or sonographer should a ground wire break. The unit, the probes, the ECG module, and all external peripherals must be tested.
  • Page 393: Chassis Leakage Current Test

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–3 Chassis Leakage Current Test (Continued) Meter Procedure Follow these steps to test the unit for leakage current. Turn the LOGIQ 500 unit OFF. Plug the unit into the meter, and the meter into the tested AC wall outlet (Refer to ILLUSTRATION 7–6).
  • Page 394 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–3 Chassis Leakage Current Test (Continued) Data Sheet for Chassis Source Leakage Current Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the table below (TABLE 7–2).
  • Page 395: Probe Leakage Current Test

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–4 Probe Leakage Current Test Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 396: Probe Leakage Current Test

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–4 Probe Leakage Current Test (Continued) Meter Procedure Using Probe Adapter The Dale 600/600E provides a method for testing probes independently from the system. The meter utilizes a probe adapter to apply a test potential commonly to all connector pins.
  • Page 397 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–4 Probe Leakage Current Test (Continued) Plug the probe adapter into the meter’s connector marked “EXTERNAL” (Refer to ILLUSTRATION 7–8). Set the meter’s “FUNCTION” switch to EXTERNAL position. Add the saline probe and the imaging area of the probe into the saline bath.
  • Page 398 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–4 Probe Leakage Current Test (Continued) Data Sheet for Transducer Source Leakage Current Follow the foregoing test procedure for every transducer. The test passes when all readings measure less than the value shown in the table below (TABLE 7–4).
  • Page 399 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–5 ECG Leakage Current Test Definition This test measures the current which would flow to ground from any of the isolated ECG leads. The meter simulates a patient who is contacted to the monitoring equipment and is grounded by touching some other grounded surface.
  • Page 400: Ecg Leakage Current Test

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–5 ECG Leakage Current Test (Continued) Meter Procedure The Dale 600/600E provides five snap type ECG buttons for testing patient leads. Snap on all patient leads to the meter and assure that the ground clip is connected to the system’s ground terminal.
  • Page 401 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–5 ECG Leakage Current Test (Continued) Set the meter’s “FUNCTION” switch to LEAD TO GROUND position to measure the patient lead to ground leak- age current. Select and test each ECG lead positions (except ALL) of the LEAD selector, testing each to the power condition combinations shown in TABLE 7–6.
  • Page 402 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–5 ECG Leakage Current Test (Continued) Data Sheet for ECG Leakage Current Follow the foregoing test procedure. The test passes when all readings measure less than the value shown in the table below (TABLE 7–7).
  • Page 403: When There's Too Much Leakage Current

    GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA 7–3–6 When There’s Too Much Leakage Current... Chassis Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part.
  • Page 404 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 1 P9030TA CHAPTER 8 – INSTALLATION FOR OPTIONS TABLE OF CONTENTS SECTION TITLE PAGE 8–1 INTRODUCTION ..............
  • Page 405 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 1 P9030TA This page is left blank intentionally. 8–2 INSTALLATION FOR OPTIONS...
  • Page 406 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 1 P9030TA 8–1 INTRODUCTION This chapter is provided to keep option installation instructions. Keep the option installation instructions on this chap- ter after completing installation for options. 8–3 INSTALLATION FOR OPTIONS...
  • Page 407 GE MEDICAL SYSTEMS LOGIQ 500 SERVICE MANUAL REV 10 P9030TA This page is left blank intentionally. 8–4 INSTALLATION FOR OPTIONS...