LV Calculation Formulas (LV SP-DISC Method) (cont'd)
Calc Mnemonic
LVAd
Left Ventricular Area,
Diastole
LVAs
Left Ventricular Area,
Systole
HR
Heart Rate
(beats/minute)
EdV
End Diastole Volume
EsV
End Systole Volume
SV
Stroke Volume
EF
Ejection Fraction
CO
Cardiac Output
SI
Stroke Volume Index
CI
Cardiac Index
(*1) where a
is the i
di
(*2) where a
i is the i
s
Reference: Journal of American Society of Echocardiolography, Vol. 2 Nov 5, 1989.
Table 4–5. LV SP-DISC Method Formulas (cont'd)
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
Calc Name
one area (by trace only) is
divided into 20 elliptical
disk segments
one area (by trace only) is
divided into 20 elliptical
disk segments
one 2 beat time interval
one distance and two
areas (by ellipse, trace or
circle)
one distance and two
areas (by ellipse, trace or
circle)
two distances and two
areas (by ellipse, trace or
circle)
two distances and two
areas (by ellipse, trace or
circle)
two distances and two
areas (by ellipse, trace or
circle) and one 2 beat time
interval
body surface area and SV
body surface area and CO
th
disc diameter of LVAd.
th
disc diameter of LVAs.
Advanced Cardiac Calculations
Input Measurements
Formula
LVAd=a1[cm ^2]
LVAs=a1[cm ^2]
HR[BPM]=120 [sec]/ 2
beat time [sec]
EdV=
(*1)
20
2
(p/4) * ((LVLd/20)
a
di)
i=1
EsV=
(*2)
20
2
(p/4) * ((LVLs/20)
a
si)
i=1
SV[ml]=EdV–EsV
EF=SV/EdVx100
CO[1/min]=SVxHR/1000
SI[ml/m^2]=SV/BSA
CI[ml/min/m^2]=CO/BSA
4–69