Typical Tests To Perform - GE LOGIQ 500 Advanced Reference Manual

Hide thumbs Also See for LOGIQ 500:
Table of Contents

Advertisement

Typical Tests to Perform

Frequency of tests
Every 3 Months
Or
Every 400 Patients
Or
If system
performance
in question
LOGIQ
500 Advanced Reference Manual
2276614–100 Rev. 0
Quality assurance measurements provide results relating to
system performance. Typically these are:
S
Axial Measurement Accuracy
S
Lateral Measurement Accuracy
S
Axial and Lateral Resolution
S
Penetration
S
Functional & Contrast Resolution
S
Gray Scale Photography.
With these tests, a performance baseline can be set at
installation with the phantom in your department. Future test
results can be compared to the baseline in order to maintain a
record of system performance trends.
Quality assurance tests are used to determine whether a
scanner is providing the same level of performance from day to
day.
The frequency of testing varies with the amount of system
usage and modes to be tested. It is recommended that the user
perform quality assurance tests at least every three months or
every 400 patient studies. Tests should also be performed when
a question about system performance exists.
A mobile system may require more frequent tests.
Image quality should also be tested immediately after the
following events:
S
Service calls
S
System upgrades/modifications
S
Dropped probe, power surge, etc.
Quality Assurance